4 Mounting
4.1 Mounting preparations
The following tools are required for dismounting the existing and
mounting the spare process component:
•
Magnetic screwdriver Torx size T10 or slot 4
•
pliers for locking ring (version axle)
•
Screwdriver, slot size 1
•
Screwdriver slot 4 (with stainless steel single chamber as well as
stainless steel and aluminium double chamber housing)
4.2 Dismounting steps previous process
component
The electronics module is located in the electronics compartment.
The below illustrations show the respective position of the electronics
compartment in a single or double chamber housing.
1
Fig. 1: Single chamber housing
1 Position of the electronics compartment
1
Fig. 2: Double chamber housing
1 Position of the electronics compartment
Proceed as follows:
1. Switch off voltage supply
2. Remove the housing lid from the electronics compartment, or
from the terminal compartment of the aluminium double chamber
housing, as the case may be
3. Disconnect the connection cables according to the operating
instructions manual of the respective sensor
4. Loosen the two holding screws with a screwdriver (Torx size T 10
or slot size 4)
Tools
Dismounting the elec-
tronics module
Summary of Contents for VEGABAR 80 Series
Page 1: ...Operating Instructions Process module VEGABAR series 80 Document ID 1013720...
Page 12: ...12 Notes Process module VEGABAR series 80 1013720 EN 220826...
Page 13: ...13 Notes Process module VEGABAR series 80 1013720 EN 220826...
Page 14: ...14 Notes Process module VEGABAR series 80 1013720 EN 220826...
Page 15: ...15 Notes Process module VEGABAR series 80 1013720 EN 220826...