9 Diagnosis, asset management and service
VEGAPULS 69 • Two-wire 4 … 20 mA/HART
47247-EN-200427
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
F261
Error in the instru-
ment settings
Error during setup
False signal suppression faulty
Error when carrying out a reset
Repeat setup
Carry out a reset
Byte 4, Bit 1 of
Byte 0 … 5
F264
Installation/Setup
error
Adjustment not within the vessel
height/measuring range
Max. measuring range of the in-
strument not sufficient
Check or correct installation and/
or parameter settings
Use an instrument with bigger
measuring range
Byte 4, Bit 2 of
Byte 0 … 5
F265
Measurement func-
tion disturbed
Sensor no longer carries out a
measurement
Operating voltage too low
Check operating voltage
Carry out a reset
Disconnect operating voltage
briefly
Byte 4, Bit 3 of
Byte 0 … 5
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
C700
Simulation active
A simulation is active
Finish simulation
Wait for the automatic end after
60 mins.
"Simulation Active"
in "Standardized
Status 0"
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
S600
Impermissible elec-
tronics temperature
Temperature of the electronics in
the non-specified range
Check ambient temperature
Insulate electronics
Byte 23, Bit 0 of
Byte 14 … 24
S601
Overfilling
Danger of vessel overfilling
Make sure that there is no fur-
ther filling
Check level in the vessel
Byte 23, Bit 1 of
Byte 14 … 24
Code
Text message
Cause
Rectification
DevSpec
State in CMD 48
M500
Error during the re-
set "delivery status"
The data could not be restored
during the reset to delivery status
Repeat reset
Load XML file with sensor data in-
to the sensor
Byte 24, Bit 0 of
Byte 14 … 24
M501
Error in the non-
active linearisation
table
Hardware error EEPROM
Exchanging the electronics
Send instrument for repair
Byte 24, Bit 1 of
Byte 14 … 24
M502
Error in the event
memory
Hardware error EEPROM
Exchanging the electronics
Send instrument for repair
Byte 24, Bit 2 of
Byte 14 … 24
Function check
Out of specification
Maintenance