Section11: System Maintenance and Care 

 
Daily Maintenance 
It is recommended that you perform the following 
maintenance procedures to help minimize bio-film 
formation in the water path of your Little Beaver 2.0 
ultrasonic scaler which could affect the water flow to 
the ultrasonic insert, and scaling performance. 
 
Start-Up  Procedures  at  the  beginning  of  the 
day: 
1.  Open  the  manual  shut-off  valve  on  the  dental 
office water supply system. 
2.  Turn  the  System  ON  using  the  Power  ON/OFF 
switch. (see illustration on page 7) Verify the Power 
Indicator Light is on. 
3. Set the Power Control knob to minimum setting. 
4. Set the Water Control knob to maximum. 
5. Hold the Handpiece (without an insert installed) 
upright  over  a  sink  or  drain.  Activate  the  Foot 
Control  and  flush  the  water  line  for  at  least  2 
minutes. 
6. Place a sterilized insert into the Handpiece and 
set  the  water  control  knob  to  your  preferred 
operating position. 
 
Between Patients: 
1.

 

Remove  ultrasonic  insert  and  handpiece  sleeve 
used, clean and sterilize. 

2.

 

Clean  and  disinfect  the  surfaces  of  the  cabinet, 
Power Cord, Handpiece cable assembly*, control 
knobs,  Foot  Control  and  cable  assembly  by 
applying  an  approved  non-immersion  type 
disinfectant  solution*  carefully  following  the 
instructions provided by the disinfectant solution 
manufacturer. To clean system, generously spray 
disinfectant solution on a clean towel and wipe all 
surfaces. Discard used towel. To disinfect system, 
generously  spray  disinfectant  on  a  clean  towel 
and wipe all surfaces. Allow disinfectant solution 
to  air  dry.  Do  not  spray  disinfectant  solution 
directly on the ultrasonic scaler. 

3.

 

Place a sterilized handpiece sleeve. Set power to 

minimum. Hold the handpiece over a sink or drain 
and flush the water line at maximum water flow 
for 30 seconds. 

4.

 

When  ready,  place  a  sterilized  insert  into  the 
handpiece. 

5.

 

Please consider use a FDA approved sheath to for 
the cable and entire handpiece or at least to cover 
the  handpiece  from  the  fastening  ring  to  cable 
when sleeve is sterilized between patients. 
 

Shut-Down Procedures at the end of the day: 
2.

 

Remove  ultrasonic  insert  and  handpiece  sleeve 
used, clean and sterilize. 

3.

 

Turn the System OFF. 

4.

 

Clean  and  disinfect  the  surfaces  of  the  cabinet, 
Power  Cord,  Handpiece  cable  assembly,  control 
knobs, Foot Control cable assembly by applying 
an  approved  non-immersion  type  disinfectant 
solution*  carefully  following  the  instructions 
provided 

by 

the 

disinfectant 

solution 

manufacturer. To clean system, generously spray 
disinfectant solution on a dean towel and wipe all 
surfaces. Discard used towel. To disinfect system, 
generously  spray  disinfectant  on  a  clean  towel 
and wipe all surfaces. Allow disinfectant solution 
to  air  dry.  Do  not  spray  disinfectant  solution 
directly on the ultrasonic scaler. 

5.

 

Close  the  manual  shut-off  valve  on  the  dental 
water supply system. 

 
*Note:  Those  Water-based  disinfection  solutions 
are preferred, due some alcohol-based disinfectant 
solutions may be harmful and may discolor plastic 
materials. 
 
*Note: The electric wire winding covered by heat 
shrinkage tube which protected by handpiece sleeve 
is  very  sensitive  to  the  disinfectant  solution  and 
water.  After  the  disinfection,  wipe  surface  of 
shrinkage tube with a slightly damp cloth and dry 
thoroughly before use. 

Summary of Contents for Lil' Beaver 2.0

Page 1: ...S 19 01 1214 UV101 V1 3 VECTOR LiTTLE BEAVER 2 0 ULTRASONIC SCALER 25 30K Instruction For Use VECTOR R D INC 6824 19TH ST 230 UNIVERSITY PLACE WA 98466 USA Web www vectorusa net User Manual...

Page 2: ...7 2 Water Line Requirements 7 3 Electrical Requirements 7 4 Unpacking the System 7 5 System Installation 7 6 Vertical Installation 7 7 Rear Panel Controls Power Connection 7 8 Water Supply Line Connec...

Page 3: ...ffice The use of an in line water filter is recommended Never operate the ultrasonic scaler without water flowing through the handpiece Grounding reliability can only be achieved when the equipment is...

Page 4: ...rms to IEC60601 1 2 2007 and IEC60601 1 1 2005 The device is CE marked corresponding to European Medical Device Directive 93 42 EEC 2 2 Supplies and Replacement Parts Contact your local Vector R D Inc...

Page 5: ...0 60Hz 1 0 A MAX 230VAC 50 60Hz 0 6 A MAX WATTAGE MAX 85 W WATER PRESSURE 20 40 psi 4 2 Dimension HEIGHT 6 5 CM WIDTH 23 0 CM DEPTH 22 0 CM WEIGHT 2 1 KG 4 3 Operation Environment AMBIENT TEMPERATURE...

Page 6: ...nd patient safety carefully follow section 11 system maintenance and care information detailed in the operating instruction accompanying your ultrasonic scaler As with high speed handpieces and other...

Page 7: ...verify there are no leaks 7 3 Electrical Requirements Refer to Section 3 Specifications 7 4 Unpacking the System Carefully unpack your Little Beaver 2 0 ultrasonic scaler and verify that all componen...

Page 8: ...ector Connect foot switch here ensure it is firmly seated Solenoid Water Inlet Power Socket Intensity Control Increase or decrease power level for operation increases power decreases power Handpiece O...

Page 9: ...ning ring and rotate the sleeve counterclockwise from handpiece tubing assembly Note Avoid rotate or twist the handpiece tubing 8 4 Foot Switch Controls and Operation The Foot Switch is a two position...

Page 10: ...NTO PLACE Activate the System Hold the handpiece over a sink or drain Check spray temperature to verify fluid is reaching the working end of the insert tip Adjust the water cooler irrigate Control kno...

Page 11: ...ant solution to air dry Do not spray disinfectant solution directly on the ultrasonic scaler 3 Place a sterilized handpiece sleeve Set power to minimum Hold the handpiece over a sink or drain and flus...

Page 12: ...or Light is not lit 1 Check that the Power switch is in the ON position and that the detachable Power Cord is fully seated in the receptacle on back of System 2 Check that the System s three prong plu...

Page 13: ...0 Ultrasonic Scaler in accordance with local and national laws Section 14 Disclaimer VECTOR R D INC considers itself responsible for the effects on safety reliability and performance of this product o...

Page 14: ...ranty period are not carried out by an authorized engineer dealer warranty will expire immediately 15 3 Warranty Exception The warranty stated herein is the sole warranty applicable to Vector R D Inc...

Page 15: ...manufacture Catalogue number Type BF Applied Part Type BF Equipment Protection against electric shock Equipment class Class II IEC601 1 double insulated The equipment complies with the requirements in...

Page 16: ...S 19 01 1214 UV101 V1 3 16...

Reviews: