4 Start up
10 / 41
progress. Depending on the camera model and current
j
parameter configuration, the actions
accomplished by the
k
command and its output at the serial interface may vary. Refer to the
corresponding command reference in appendix 11.1 for details.
The following table shows an overview of the consumed time by the command “
k=0
”:
Description
Time
Max. Time
Typical time for a IRC-320
Close shutter (optional)
4 * T
100ms
0ms
Save one image
3 * T
75ms
75ms
Open shutter (optional)
0 * T
0ms
0ms
Choose data set
N1 * 2 * T
1600ms
800ms (N1=16)
Close shutter
5 * T
125ms
125ms
Integrate images
(N2 + 1) * T
1625ms
1625ms
Open shutter
0 * T
0ms
0ms
Calculate statistics
2 * T
50ms
50ms
At all
3,58s
2,68s
T = Time for one frame-out (IRC-320: 25ms)
N1 = number of data sets (max. 32)
N2 = number of integrated images (max. 64)
VDS Vosskühler GmbH
IRC-320