background image

32

33

Ώ

Analyzer checks the conditions of the inserted test device. If the test device 

has already been used, E-1 error pops up.

ΐ

$IWHUFRQࢉUPLQJWKDWWKHWHVWGHYLFHFDQEHXVHGZLWK%,2127(9WKH

samples should be prepared according to the instruction for each test and 

applied to the sample well of the test device. 

Caution

 

Press the “Test Start” immediately after applying the 

sample to the test device.

Α

$QDO\]HUFKHFNVWKHDSSOLHGVDPSOHȆVFRQGLWLRQVDQGࢊRZ

Β

 When the sample checking is completed, the test device is mounted inside 

the Analyzer.

Γ

 When the mounting is completed, you can start a new test procedure.

Summary of Contents for V2400

Page 1: ...V2400 FLUORESCENT IMMUNOASSAY SYSTEMS USER MANUAL FOR VETERINARY USE ONLY FLUORESCENT IMMUNOASSAY SYSTEMS...

Page 2: ...ote co kr Thank you for your purchase of the BIONOTE V2400 This User manual contains all information about the Analyzer Please read this User manual carefully and familiarize yourself with the require...

Page 3: ...apter 03 Log On and Setting Operation of Analyzer 19 Chapter 04 Work Place Performing a test Reviewing data 30 Chapter 05 Calibration Calibration Set Test 38 Chapter 06 Quality Control Quality Control...

Page 4: ...cking Analyzing Insert Device CAL 1 Insert Device Device Checking Insert Device CAL 2 Apply Sample Apply Sample Device Checking Result Work Place Start Input Operator ID Device Checking Sample Checkin...

Page 5: ...Note To indicates that the analyzer is fragile and you need to handle it with care Batch code Indicate the lot number of the system Crossed out wheeled bin Discard it separately from other household w...

Page 6: ...Do not use test devices which have expired To prevent electric shock Ensure the Analyzer is completely grounded Do not connect the Analyzer to an ungrounded outlet When using a power board use a groun...

Page 7: ...ime after samples are dispensed 2QFH D UHDFWLRQ LV FRPSOHWHG XRUHVFHQFH G H RQ WKH FRQWURO OLQH DQG WHVW OLQH ZLOO UH HFW WKH OLJKW IURP DQG WKLV UH HFWHG YDOXH ZLOO EH PHDVXUHG Then the measured valu...

Page 8: ...H VXLWDEOH IRU DQDO VHV GL HUV LQ HDFK WHVW RU HDFK WHVW refer to the instruction manual of the test device that you will use for analyses 7R SHUIRUP D WHVW ZLWK WKH QDO HU SODFH WKH QDO HU RQ D DW VX...

Page 9: ...protect printer paper E Printer cover open button Used to open printer cover F Mini USB 8VHG WR XSJUDGH UPZDUH E FRQQHFWLQJ ZLWK 3 G LAN Used for communication using LAN Cable H AC power connection U...

Page 10: ...nnector on the back of the Analyzer and to the outlet 1 3 Once the connection is complete the analyzer is ready for use Step 2 Analyzer Log On 2 1 Press log on button 2 2 The initial log on ID is admi...

Page 11: ...ting You can set the number of result sheet to 1 sheet or 2 sheets Once press Auto Print the test result will be printed automatically after all 3 Language Setting Press a down arrow to select the des...

Page 12: ...een Press OK to reboot the analyzer f Make sure that the latest S W version is displayed in the lower left corner of the screen F W Update Items needed for F W update BIONOTE V2400 Analyzer LUPZDUH XS...

Page 13: ...status and log of each item Maintenance Area Tested items are displayed Status Area QC scheduled date for tested item is displayed The QC period can be set from due date in Area above Due Date Area Y...

Page 14: ...f errors that occurred are displayed The number of times of calibrations and Q C for each test is displayed Period setting Press button to set the period that you want to check Caution Top Clear Delet...

Page 15: ...you wish to modify and then press OK To delete the Operator ID select the ID select the ID press Delete then press OK Network Menu 1 On the Network tab you can set the HIS LIS and Analyzer network Fo...

Page 16: ...est Status and Result Review 1 Run test Menu Run test is a mode for testing the samples from patients After entering patient information in Patient ID press to move to the next step Patient ID can be...

Page 17: ...uld be prepared according to the instruction for each test and applied to the sample well of the test device Caution Press the Test Start immediately after applying the sample to the test device QDO H...

Page 18: ...ve If the result of quantitative test is out of normal range Black section is shown in the following cases If the test result is Invalid Green section is shown in the following cases If the test is in...

Page 19: ...see the tests in order from the most recent ones If you select a test from the list in Area B you can see the test details Test details window for quantitative test Example Canine CRP Test details win...

Page 20: ...e of the Analyzer and test device Caution Calibration Set test is independent from QC test and cannot replace QC test Calibration mode consists of Run Calibration and Calibration Review 1 Run Calibrat...

Page 21: ...is completed you can check the result screen Calibration check OK screen Calibration check FAIL screen 2 Calibration Review Menu The result of calibration can be checked in Calibration Review You can...

Page 22: ...ce of newly opened package When Analyzer is dropped When the result does not match the expected result When the result is lower or higher than expected even after repeated tests When you want to check...

Page 23: ...device that matches the quality control material information entered on the Insert Device screen The Analyzer checks the conditions of inserted test device If the test device has already been used an...

Page 24: ...Insert the test device to test on the Insert Device screen Analyzer checks the conditions of inserted test device If the test device has already been used an E 1 error pops up KHQ WKH YHUL FDWLRQ LV F...

Page 25: ...result of QC in QC Result review a The number of times that QC was performed average standard deviation FRH FLHQW RI YDULDWLRQ DQG WKH UHVXOW RI s 6 DUH VKRZQ RU TXDOLWDWLYH items only the number of t...

Page 26: ...down list e Deletes the entire QC result If you select from the list you can check the test details 4 Monthly Review Menu In Monthly Review you can check the monthly value of quantitative test If you...

Page 27: ...ing case is designed to store the supplied components and to protect the Analyzer 7KH QDO HU PXVW EH VWRUHG DQG WUDQVSRUWHG DW a a DQG 5 The V2400 Analyzer was designed to meet safety standards Do not...

Page 28: ...eeded the expiry date E04 Temperature error The ambient temperature deviates from the operating temperature range of the test device Solution Move the Analyzer and test device into proper temperature...

Page 29: ...se of the BIONOTE V2400 Analyzer in accordance with local regulations for disposal of electrical and electronic equipment Waste Electrical Electronic Equipment WEEE regulation corresponds to the Europ...

Page 30: ...MEMO MEMO...

Page 31: ......

Reviews: