Measurement specification:
The total measurement time for a full measurement cycle is less than 3
minutes, as opposed to a single measurement time of about 1.5 minutes.
Single results are not displayed.
Based on data analysis result, a 4
th
measurement may be applied.
2.4 Irregular
h
eartbeat
d
etection (IHD)
The appearance of the symbol
indicates that certain pulse irregularity is
detected during measurement. This may cause results to vary from your normal
blood pressure levels. In general, this is not a cause for concern. However, if the
symbol appears more frequently than usual, it is recommended that you consult
with your healthcare provider.
If SmartCheck™ measurement mode is used, irregular heartbeat detected during
measurement will not affect results (see section 2.3 for more information).
Note:
The device detects early stage pulse irregularities. It is not a replacement
for cardiac examination.
2.5 Traffic
l
ight
i
ndication
The colored bars on the left edge of the display indicate the range within which
the measured blood pressure values lie. Depending on the height (or number) of
the bars, the measured values are either within the normal (green),
elevated/hypertension stage 1 (yellow) or hypertension stage 2 (orange/red)
range. The classification corresponds to the 4 levels as defined by the American
Heart Association (AHA). Please refer to the table in section 2.2 to learn more.
In addition, the 3-color backlight display shows different color according to the
measured blood pressure values. Readings in the classification range of 1 is shown
in green backlight, classification ranges of 2, 3 in yellow backlight and
classification ranges of 4, 5, 6 in red backlight.