Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
69 / 188
COPYRIGHT© 2018 VATECH
.
ALL RIGHTS RESERVED
4.2. Performing CBCT Rotator Position Alignment
1) Run Console Software and click CBCT to go into CBCT mode.
2) Set the parameters as shown below and click CONFIRM.
3) Click READY when enabled and acquire the image according to the Console Software
guidance.
4) After the image acquisition is complete, exit the capture software and run VAKCAP.exe.
VAKCAP.exe path: C:\VCaptureSW\Acquisition\CBCT\WidePano
5) Click the Capture/Recon tab and click V Slc to open View 16.
6) Select C or R in the M box and select a slice file that shows the cross line clearly