6) Click the
Soft Calib
tab and type
[SPM_ARCT]
. Then click Send to initiate the system
into CT mode.
7) When the equipment stops operating, type
[SPM_CSON]
and click
Send
to supply
power to the CT Sensor.
8) Click the
Soft Calib
tab, wait 20 seconds and click
Dark
to achieve the Dark Calibration data
file.
9) Select
184x686d.raw
and click
V
to open View16.