PaX-i Plus / Insight (Model: PCH-30CS) Installation Manual
Determining the Height
Calculate the adjustment bolt position by using the following formula.
H
screw
height
=H
ceiling
-d, where d is the minimum distance, 30 cm (12") between the
ceiling and the top of the equipment (when the equipment is fully loaded).
(e.g. If d = 30 cm, H
ceiling
= 250 cm (99”), H
screw height
= H
ceiling
-d = 250-30= 220 cm.
Accordingly, set the height to this level by using the height adjustment bolt.