background image

28

Step 2

Select the tooth position to capture. A blue rectangle will be drawn on the tooth box.

Step 3

Click the ‘SENSOR (                )’ button or ‘Capture (                  )’ button.

Step 4

Position the sensor correctly.

https://stomshop.pro

Summary of Contents for EzSensor HD

Page 1: ...Installation User Manual IOS U20VF IOS U15VF IOS U10VF IOS U20IF IOS U15IF IOS U10IF https stomshop pro...

Page 2: ...lder 14 CHAPTER 4 IMAGE ACQUISITION BY USING EASYDENT 15 4 1 Installation of EasyDent 15 4 2 Image Acquisition Overview 20 4 3 Running the EasyDent 21 4 3 1 Patient Registration and Device Selection 2...

Page 3: ...5 6 2 1 Test List 56 6 2 2 Connection 56 6 2 3 Cable 57 6 2 4 Resolution 57 6 3 Care and Cleaning 58 6 4 Precautions 58 6 5 Product complaint 59 APPENDIX 62 A 1 X ray Exposure Guide 62 A 2 Error Messa...

Page 4: ...manual depending on individual Product specifications 1 2 Conventions and Symbols 1 Convention The following symbols are used throughout this manual to provide instructions on the effective use of thi...

Page 5: ...C European Communities concerning medical devices 4 Waste Electrical and Electronic Equipment 5 Manufacturer 6 Date of manufacture 7 Serial number 8 Authorized representative in the European Community...

Page 6: ...r further instructions on how to proceed To ensure the correct usage of the EzSensor HD device in a clinical environment for which the intended purposes correspond to its design and application only d...

Page 7: ...60 seconds After immersing the sensor in disinfectanct use a dry cloth to wipe the sensor Use of other disinfectant may cause discoloration but will not cause dysfunction of the sensor Explosion safe...

Page 8: ...eturn the Product to the seller on a one to one basis as long as he or she is buying a new one of equivalent type and that fulfils the same functions as the old one Disposing of the Product separately...

Page 9: ...0 1 44 x 1 00 x 0 19 inch 36 8 x 25 4 x 4 8 mm Size 1 5 1 55 x 1 15 x 0 19 inch 39 5 x 29 2 x 4 8mm Size 2 0 1 69 x 1 28 x 0 19 inch 42 9 x 31 3 x 4 8 mm Pixel Pitch High Resolution mode 0 0148 mm No...

Page 10: ...700 to 1060 hPa EU classification Medical Devices 93 42 EEC as a class IIa Protection against shock Type B applied part Protection against matter Water IP68 The Product has to be installed transported...

Page 11: ...marks 1 Sensor Module 2 7m USB cable is all in one 2 Wrap Hygienic Sleeves 3 Holder for Sensor 4 Silicon cover 5 S W Installation CD or USB EasyDent or EzDent i 6 Driver Installation CD USB Device Dri...

Page 12: ...ing System Window 34 64bit XP SP1 Vista 7 8 8 1 We cannot guarantee that Easydent EzDent i will work properly with an unregistered copy of Microsoft Windows Therefore you should use a registered genui...

Page 13: ...driver is also installed during this step Step 1 Insert the S W Installation CD in the CD ROM Setup should start automatically If it does not click Start Run and type Homedirctory Console setup exe St...

Page 14: ...then click the Next button Step 4 The Install shield Wizard will start configuring the installation parameters The Install shield wizard will Copy the calibration files To your work station PC Step 5...

Page 15: ...PC directly Be sure to connect the USB port on backside for accurate operation Do not connect the EzSensor HD and USB PC Interface cable to your computer until you have successfully installed the setu...

Page 16: ...System Device Manager Windows XP Settings Control Panel System Hardware Device manager Select VH EzSensor N x x located under Imaging Devices You should see the message This device is working properly...

Page 17: ...in use When choosing where to install the EzSensor HD locate an area that offers easy access and visibility during patient examinations Position the holder on a stable flat surface Using the holes at...

Page 18: ...t will be installed Alternatively software which supports the Twain interface can be used instead of EasyDent Step 1 Insert the S W installation CD in the CD ROM drive Homedirectory EasyDent setup exe...

Page 19: ...asic Installs the basic version of EasyDent V4 CD Publishing Installs the basic version of EasyDent V4 along with CD Publishing capabilities optional For the EasyDent server select all items For PC be...

Page 20: ...ton EasyDent Viewer is the minimum requirement for EzSensor HD use Step 6 Now you are ready to install If you would like to review any of your installation settings click Back To proceed with the inst...

Page 21: ...ntV4 Step 8 The program compatibility and Windows Firewall alarm messages are shown Click Run program button on the program compatibility message And click Allow access button on the windows security...

Page 22: ...Step 9 MSDE Microsoft SQL server Desktop Engine is installed automatically Close the SQL Server Service Manager Step 10 Select No I will restart my computer later and then click Finish https stomshop...

Page 23: ...e sensor must face the X ray source Note that the receptor side is marked with a label for ease of recognition The use of the sensor positioning aid is recommended to guarantee that the sensor is para...

Page 24: ...st be fully aware of the explanations before capturing image 4 3 1 Patient Registration and Device Selection Step 1 Turn on the PC Run EasyDent4 Viewer Click the Patient button to register a new patie...

Page 25: ...22 Step 2 The new patient registration window will appear You are asked to register the new patient and click Add https stomshop pro...

Page 26: ...information Click Help Intra Oral Sensor Setting Select Device on the menu bar Select your capture device At this time you are asked to select the device and image quality The sensor model and image p...

Page 27: ...the Layouts tab Select your favorite layout from the Layout template list The Layout Template can be customized Please refer to the EasyDent manual Step 3 Select the Single from the Shoot Type list ht...

Page 28: ...25 Step 4 Select the tooth position to capture The tooth box is selected in blue Step 5 Click the SENSOR button or Capture button Step 6 Position the sensor correctly https stomshop pro...

Page 29: ...the correct position expose the X ray after the Please expose X ray message appears The message Optimizing Image Please wait appears while the image is being optimized The image will appear after opt...

Page 30: ...hot You can capture multiple images continuously until you click stop Click and drag the image to its appropriate position after stopping capture Step 1 Select the Multiple from the Shoot Type list ht...

Page 31: ...28 Step 2 Select the tooth position to capture A blue rectangle will be drawn on the tooth box Step 3 Click the SENSOR button or Capture button Step 4 Position the sensor correctly https stomshop pro...

Page 32: ...29 Step 5 Expose the X ray after the Please expose X ray message appears The message Optimizing Image Please wait appears while the image is being optimized https stomshop pro...

Page 33: ...ray when the Press the X ray shot button message appears To finish double click on the Press the X ray shot button message box And then the image will appear after optimization is complete Step 7 Cli...

Page 34: ...e images appear automatically in the correct positions Step 1 Select your favorite sequence from the Shoot Type list Red numbers represent the capture order The Sequence shot can be customized Please...

Page 35: ...the order the images appear automatically in the correct positions Expose the X ray after the Please expose X ray message appears The message Optimizing Image Please wait appears while the image is be...

Page 36: ...message box And then according to the capturing order the images appear automatically in the correct positions All images are saved in the Database server automatically By double clicking on the image...

Page 37: ...oftware which supports the Twain interface can be used instead of EzDent i Step 1 Open Setup exe from Setup folder to install EzDent i Sever or Client for Windows Step 2 The window for selecting insta...

Page 38: ...DB click the EzServer button Step 4 To continue click the Next button followed by the install button to proceed When the installation of EzServer is complete the other components Consult Data and Imp...

Page 39: ...s select complete click the Next button to proceed Step 6 In case that Custom in implant DB is selected Select the name of manufacturers you want to install and click the Next button to go forward If...

Page 40: ...pile setup exe and click the button EzDent i The License Agreement window will be shown If the conditions are acceptable select I accept the terms of the license agreement and click Next button https...

Page 41: ...38 Step 8 Type in the company and user name and click the Next button Step 9 After specifying the location to install the program click the Next button to go forward https stomshop pro...

Page 42: ...complete select Yes I want to restart my computer now and click the Finish button Then the PC will be restarted It is strongly recommended to restart the PC to apply the changes occurring in the syste...

Page 43: ...p 1 Connect the USB Dongle Key to the PC and open VTNL exe Opening VTNL To open VTNL program install ADB exe file first If the executable file is not installed the installation process of ADB program...

Page 44: ...have not activated a license previously Select the dongle key first item and then click the Next button Step 3 When the following VTLicense Plug in Option Window appears select the license key from t...

Page 45: ...been activated click on the Finish button to complete your activation EzDent i will then run automatically Once the license has been activated the user s PC will appear on the list of Certifications...

Page 46: ...Dent i user information and license key provided And then click the Next button You must enter more than 3 characters to activate the Next button For activating license using the license key an intern...

Page 47: ...on Wizard window will appear including the remaining days Click the Finish button to end the activation Restart EzDent i after the license activation The user can check the information of the license...

Page 48: ...nt i Server or Client for Windows Step 2 After selecting the install modes Click the Consult License Activator button Consult DB is for the users who have a Consult Premium License After activating Co...

Page 49: ...he Consult Data License Activator window Click the Activate button Step 4 Select the second option from the VTLicense Activation Wizard window For activating license using the license key an internet...

Page 50: ...d then click the Next button You must enter more than 3 characters to activate the Next button Step 6 Upon successful activation the VTLicense Activation Wizard window will appear including the remain...

Page 51: ...must face the X ray source Note that the receptor side is marked with a label for ease of recognition The use of the sensor positioning aid is recommended to guarantee that the sensor is parallel to t...

Page 52: ...5 6 1 Patient Registration Step 1 Turn on the PC Run EzDent i Click the Patient button to register a new patient Step 2 The Add Patient registration window will appear Register the new patient and cl...

Page 53: ...cquisition Tab allows the user to capture store images with the IO Sensor By clicking the logo EzDent i as below you can use normal resolution mode Normal Resolution mode High Resolution mode Select a...

Page 54: ...Once the IO sensor is connected the layout will change depending on the selected device Step 3 Click the Ready button after positioning the patient If the IO Sensor equipment is properly connected the...

Page 55: ...d the image will be acquired The process of acquiring the image is complete By double clicking the workspace from the ACQUISITION tab EzDent i will be ready for image acquisition The Ready button prov...

Page 56: ...to select more than 2 images hold the Ctrl key on the keyboard and click the desired images from the thumbnails To change the way of storing the acquired images change the options in settings With th...

Page 57: ...art number of the patient in the Patient Search window and click the Search button or press the Enter key The search result will appear Step 2 Double Click the selected Patient List The images will be...

Page 58: ...t your local Product distributor for further instructions on how to proceed 6 2 Periodic Maintenance Periodic maintenance should be performed as necessary and in accordance with monitored frequency in...

Page 59: ...Check the normal operation of PC when the sensor is connected to the PC Procedure 1 Connect the EzSensor HD USB Connector to the USB Port 2 Confirmation of Driver installation at the Device Manager Me...

Page 60: ...amage of the cable s rubber tube They should not be stepped on nor bent and pressed under table legs 2 Check that all cables connected to the EzSensor HD are undamaged 6 2 4 Resolution Object Verify t...

Page 61: ...cotton swab dipped in one of the cleaning solutions listed above Gently wipe the surface end to end in straight lines without applying any pressure Make sure the liquid does not penetrate the Product...

Page 62: ...ndle such issues If the Product may have caused or contributed to a serious injury of a patient your distributor should immediately notify the manufacturer by telephone fax or written correspondence T...

Page 63: ...on of the Product in a field or in an environment for which such the Product was not designed or contemplated v use of any parts or material not provided by MANUFACTURER for warranty service vi the th...

Page 64: ...f the use or inability to use the Product even if the Buyer has been advised of the possibility of such damages If the Buyer fails to pay any amounts due to the Seller whether related to the Products...

Page 65: ...sing reduces the difference between image qualities of different doses Users can adjust brightness and contrast in the option menu The recommended exposure dose is from 300 Gy to 600 Gy when measuring...

Page 66: ...posure condition can be changed depending on the age gender and bone density of the patient in case of Pediatric X ray exposure condition can be changed by expert s judge For further information pleas...

Page 67: ...omer Service 6 Data communication error Solution Re connect the USB PC cable 7 Canceled image capturing This means that the user canceled image capture Please try again 8 Cannot find dark frame Soluti...

Page 68: ...he X ray source If the problem persists call for technical assistance 12 Cannot load EzSensor dll Solution Please re install the acquisition software 13 Require EzSensor dll was damaged Solution Pleas...

Page 69: ...m persists please contact your local Product distributor Table 5 Troubleshooting Table Item Description Corrective Measure 1 A PID 2XXX NO 0 Check Connection error message is displayed Unplug the USB...

Page 70: ...onment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tiles If floors are covered with synthetic material rela...

Page 71: ...ccur in the vicinity of equipment marked with the following symbol Note 1 U is the A C mains voltage prior to application of the test level Note2 At 80 MHz and 800 MHz the higher frequency range appli...

Page 72: ...rding to frequency of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 70 3 70 7 37 100 11 70 11 70 23 30 For transmitter...

Page 73: ...manual If you do not properly set the Product which in turn causes the Product to malfunction or fail we cannot guarantee any responsibility Rayence Co Ltd Web Site4www rayence com 14 Samsung 1 ro 1 g...

Page 74: ...0 Document No R USM 712 Release Version 1 4 HD Dated Jul 8 2016 Distributed by Vatech 13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do Korea https stomshop pro...