5. Troubleshooting
EzRay Chair (VEX-S300C) User Manual
37
EN
G
LISH
5. Troubleshooting
In instances of abnormal operation, alarm/error messages will be displayed on the
Control Panel. If a problem persists, please request assistance from the customer
support information services.
Alarm/Error Messages
A.0X:
A problem occurred, and the system performs the
correction automatically. This alarm clears after the correction
is completed.
E.0X:
An error occurred. Turn the power off, and then turn it
back on. If the error persists, contact your Service
Representative.
Error Code
Check Parameter
Description
E.02
X-ray Generator
Error related to X-ray exposure is not
possible to exposure X-ray in the state while
"E.02", "E.03", "E.04", "E.05" where power is
maintained. After X-ray exposure related
error occurs, when the equipment is turned
off and turned on, X-ray exposure is usually
performed.
E.03
E.04
A.06
X-ray Generator
The system needs cooling time due to
continuous operation. (Duty Cycle:
1:60
)
This alarm clears when the system
temperature goes down to normal.
A.07
System
The Remote Exposure Switch has been held
down. Release the switch.
A.08
User
The exposure button has been pressed and
released before the X-ray exposure finished
with the default exposure time. Press and
hold the exposure button for the duration of
the exposure time.
Summary of Contents for EzRay Chair
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice ii EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 20: ...2 Safety Instructions 12 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 44: ...4 Operation 36 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 48: ...6 User Maintenance 40 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 54: ...9 Product Specifications 46 EzRay Chair VEX S300C User Manual...
Page 55: ...9 Product Specifications EzRay Chair VEX S300C User Manual 47 ENGLISH...
Page 56: ...9 Product Specifications 48 EzRay Chair VEX S300C User Manual...
Page 60: ...9 Product Specifications 52 EzRay Chair VEX S300C User Manual 4 Tube Dimensions mm...
Page 80: ...0 Appendix 72 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......