4. Operation
24
EzRay Chair (VEX-S300C) User Manual
3. After tooth type selection, the patient type should be selected. When the
Adult/Child selection area flickers, turn the jog dial to select the patient type. To
see the Control Panel after selection, refer to the figure below.
After patient type selection
Patient Type
Symbol
Type
Adult
Child
After the tooth type and patient type are selected, the
exposure time is automatically displayed.
4. If you want to change the exposure time, turn the jog dial to adjust the exposure
time from 0.05 to 0.5 s. (increments: 0.01 s)
If you press the jog dial after adjusting the exposure time in
Manual Mode, the exposure time is returned to the default
setting.
To save the exposure time as default in Manual Mode, press
and hold the jog dial for about 3 seconds.
Summary of Contents for EzRay Chair
Page 1: ......
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...Notice ii EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 20: ...2 Safety Instructions 12 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 44: ...4 Operation 36 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 48: ...6 User Maintenance 40 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 54: ...9 Product Specifications 46 EzRay Chair VEX S300C User Manual...
Page 55: ...9 Product Specifications EzRay Chair VEX S300C User Manual 47 ENGLISH...
Page 56: ...9 Product Specifications 48 EzRay Chair VEX S300C User Manual...
Page 60: ...9 Product Specifications 52 EzRay Chair VEX S300C User Manual 4 Tube Dimensions mm...
Page 80: ...0 Appendix 72 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......