background image

   

4.3

 Touch the DIM switch on the control panel; adjust the dimmer to the required dim   

 

point. Touch the DIM switch again; the lights should be fully 

on

.  Touch the OFF  switch  and 

 

the lights should go

 off

.  Touch the OFF switch again the lights should  

come   on. 

 

Depending if your exam room is incandescent lighting only or a combination of 

 

incandescent/fluorescent lighting, the following should happen:  the incandescent light  

will 

 

be at some preset DIM level that you have set and will remain  

at that level. In the off 

 

mode the Z9000 will turn both incandescent and fluorescent lights off.  In the DIM mode  only 

 

the fluorescent lights will be turned off automatically. 

 
  

4.4 

An individual well can be set to the following when the instrument is taken out of the   

 

      well. 

 
 

 

4.4.1

 room lights off 

 

 

 

4.4.2

 room lights dim 

 
 

 

4.4.3

 no lighting change 

 
 

 

4.4.4

 fixation or muscle light on 

 

4.5

 When using the indirect ophthalmoscope, room lights can be set to off or dim when    the 

indirect is lifted.  

 
 

4.6

 When the slit lamp control switch (optional) is turned on it will supply power to the  

 

      slit lamp and dim the room lights or turn them off.  Maximum load: 60 watts. 

 
 

4.7

 The Z9000 Ophthalmic Control System can also be programmed to dim the room  

 

      lights automatically when the projector button is pressed. 

 
 

4.8

 The auxiliary button activates a 115VAC circuit that can be used for any device.  This 

 

      is wired separately from the power box mounted in the wall.  Maximum load: 60  

 

      watts. 

 
 

4.9

 The lamp switch primarily switches 115VAC receptacle that is mounted near the  

 

       instrument stand to turn the lamp on the stand on or off. Maximum load: 60 watts.  

 
 

4.10

 The fixation switch is a low voltage output (12 VDC 40mA) used to manually turn   

 

        on the fixation light. 

 
 

4.11

 The muscle light is low voltage output (12 VDC 40mA) used to turn on the eye level 

 

        muscle light. 

 

5.0 ADJUSMENTS: Programming the Ophthalmic Control System 
 
 5.1

 The unit is programmed by dipswitches on the under side of the unit.  Dipswitches are 

 

      tiny switches that can be turned on and off using a ballpoint pen. If the switches are   

 

      turned on the following automatic functions occur: 

 
 

 

 

 
 

Summary of Contents for Z9000 Series

Page 1: ...620 Parkway Broomall PA 19008 610 356 3995 www varitronics com Varitronics Inc Z9000 Series Manual Copyright 2002 Varitronics Inc...

Page 2: ...rly connected Take the proper steps to turn off the power leaving all personnel and patients in safe condition if anything is found wrong with the system See Appendix C Be sure to observe the followin...

Page 3: ...he safest systems on the market today It has been engineered with many safety features to prevent shock and fire hazard In addition there are no electromagnetic interference concerns Be sure to observ...

Page 4: ...5 ADJUSTMENTS Programming the Z9000 Ophthalmic Control System 6 MAINTENECE 7 TROUBLESHOOTING 8 TERMS OF TRANSPORTATION APPENDIX A Z9000 PC Board Schematic APPENDIX B Z9000 Wiring Diagrams APPENDIX C...

Page 5: ...has configurable jumpers J3 J6 to provide incandescent or incandescent with fluorescent room lighting control See Appendix A for jumper settings 1 2 Termination Box A 1 2 1 Termination Box A requires...

Page 6: ...wer box will vary depending on which model is purchased These units are meant to work with 115 VAC only 2 1 4 Have the doctor or architect check all intended locations before installing ATTENTION Inst...

Page 7: ...d neutral voltage lines The fused circuit must be on the hot voltage line Please see electrical schematics for exact details Determine the location for Box B Box B the connections for low voltage func...

Page 8: ...ixtures These lamps are to be wired in parallel Wires shall run to Power Box A wire 7 Maximum load 300 watts 3 9 FINAL NOTES 3 9 1 For final installation please refer to Varitronics wiring schematic 3...

Page 9: ...the indirect is lifted 4 6 When the slit lamp control switch optional is turned on it will supply power to the slit lamp and dim the room lights or turn them off Maximum load 60 watts 4 7 The Z9000 Op...

Page 10: ...ndirect Ophthalmoscope Room lights dim 2 Indirect Ophthalmoscope Room lights off 3 Projector Room lights dim default to on 4 Not used 5 Not used 6 Not used 6 MAINTINENCE 6 1 There are no periodic main...

Page 11: ...ll only change the three external fuses A qualified technician should change any internal fuses See Appendix A 6 3 1 First turn the power switch off and disconnect the main AC plug on the power supply...

Page 12: ...battery date code If beyond date replace batteries 7 1 5 Problem Auxiliary Lamp Slit Lamp not working Solution Replace Fuse marked F1 Refer to Appendix A 7 1 6 Problem System is completely non functio...

Page 13: ...APPENDIX A...

Page 14: ...switch in Fused hot return Slit Lamp Pin J1 J2 1 2 3 4 5 6 7 8 9 10 11 12 0V On Off Projector Muscle Fixation Aux1 Lamp Dimmer Child Aux2 5 Volts On Off Projector Muscle Fixation Aux1 Lamp Dimmer Pwr...

Page 15: ...APPENDIX B...

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ...APPENDIX C...

Page 20: ...ASB 4 ASB INTERIOR LABELS WARNING For continued protection against fire hazard replace with same type and rating of fuse only MODEL Z9000 Desktop 1P 2P Full Load 700 Watts 115 VAC 60 Hz Only Varitron...

Page 21: ...APPENDIX D...

Page 22: ...Specification Sheet Z9000...

Page 23: ...Specification Sheet Z9000 1P...

Page 24: ...Specification Sheet Z9001 2P...

Page 25: ...Specification Sheet FW9001 2P...

Page 26: ...Specification Sheet Z9001 2P...

Page 27: ...Specification Sheet Z9001 2P...

Page 28: ...APPENDIX E...

Page 29: ...PHOTOS OF Z9000 SERIES OPHTHALMIC CONTROL SYSTEMS...

Page 30: ...PHOTOS OF ACCESSORIES...

Page 31: ...PHOTOS OF ACCESSORIES...

Reviews: