background image

 

1. NOMENCLATURE 

 

1.1 Printed Circuit Board Overview:

 

 

 

1.1.1

 The Z9000 PC Board provides discreet logic control over both room lighting and 

low voltage auxiliary devices. The board also recharges battery operated handheld 
ophthalmic devices (provided by others). 

 

1.1.2

 The board has three user changeable fuses and one internal fuse that is only to be 

accessed by a factory authorized technician.  See 

Appendix A

 for fuse requirements. 

 

 

      1.1.3

 There are two dipswitches on the bottom of the board to supply configurable  

 

      automatic functions. To change the factory default settings, refer to

  

 

 

          

 

      ADJUSTMENTS:

 Programming the Z9000 Ophthalmic Control System. 

 

 

1.1.4 

The board has configurable jumpers (J3, J6) to provide incandescent or 

incandescent with fluorescent room lighting control. See 

Appendix A

 for jumper settings. 

 

 

  1.2 Termination Box A 

 

1.2.1

 Termination Box A requires a standard four gang back box (provided by others). 

 

 

1.2.2

 This box provides 115VAC to interconnect room lights and other high voltage 

 accessories. 

 

1.2.3

 Refer to 

Appendix B

 for wiring diagrams. 

 
 

 1.3 Termination Box B 

 

1.3.1

 Termination Box B requires a standard single gang back box (provided by 

 others). 
 

1.3.2

 This box provides low voltage to interconnect remote accessories. 

 

1.3.3

 Refer to 

Appendix B

 for wiring diagrams. 

 
 

 1.4 Remote Accessories 

 

1.4.1

 Fixation, Muscle, and Indirect that terminate through Box B (See 

Appendix B

.) are 

low voltage devices. 

 
 

      

1.4.2

 Slit Lamp is terminated through Box A and Slit Lamp extension plate (See  

 

       

Appendix B

.) and is 115VAC device. 

 

 

 
 
 
 

Summary of Contents for Z9000 Series

Page 1: ...620 Parkway Broomall PA 19008 610 356 3995 www varitronics com Varitronics Inc Z9000 Series Manual Copyright 2002 Varitronics Inc...

Page 2: ...rly connected Take the proper steps to turn off the power leaving all personnel and patients in safe condition if anything is found wrong with the system See Appendix C Be sure to observe the followin...

Page 3: ...he safest systems on the market today It has been engineered with many safety features to prevent shock and fire hazard In addition there are no electromagnetic interference concerns Be sure to observ...

Page 4: ...5 ADJUSTMENTS Programming the Z9000 Ophthalmic Control System 6 MAINTENECE 7 TROUBLESHOOTING 8 TERMS OF TRANSPORTATION APPENDIX A Z9000 PC Board Schematic APPENDIX B Z9000 Wiring Diagrams APPENDIX C...

Page 5: ...has configurable jumpers J3 J6 to provide incandescent or incandescent with fluorescent room lighting control See Appendix A for jumper settings 1 2 Termination Box A 1 2 1 Termination Box A requires...

Page 6: ...wer box will vary depending on which model is purchased These units are meant to work with 115 VAC only 2 1 4 Have the doctor or architect check all intended locations before installing ATTENTION Inst...

Page 7: ...d neutral voltage lines The fused circuit must be on the hot voltage line Please see electrical schematics for exact details Determine the location for Box B Box B the connections for low voltage func...

Page 8: ...ixtures These lamps are to be wired in parallel Wires shall run to Power Box A wire 7 Maximum load 300 watts 3 9 FINAL NOTES 3 9 1 For final installation please refer to Varitronics wiring schematic 3...

Page 9: ...the indirect is lifted 4 6 When the slit lamp control switch optional is turned on it will supply power to the slit lamp and dim the room lights or turn them off Maximum load 60 watts 4 7 The Z9000 Op...

Page 10: ...ndirect Ophthalmoscope Room lights dim 2 Indirect Ophthalmoscope Room lights off 3 Projector Room lights dim default to on 4 Not used 5 Not used 6 Not used 6 MAINTINENCE 6 1 There are no periodic main...

Page 11: ...ll only change the three external fuses A qualified technician should change any internal fuses See Appendix A 6 3 1 First turn the power switch off and disconnect the main AC plug on the power supply...

Page 12: ...battery date code If beyond date replace batteries 7 1 5 Problem Auxiliary Lamp Slit Lamp not working Solution Replace Fuse marked F1 Refer to Appendix A 7 1 6 Problem System is completely non functio...

Page 13: ...APPENDIX A...

Page 14: ...switch in Fused hot return Slit Lamp Pin J1 J2 1 2 3 4 5 6 7 8 9 10 11 12 0V On Off Projector Muscle Fixation Aux1 Lamp Dimmer Child Aux2 5 Volts On Off Projector Muscle Fixation Aux1 Lamp Dimmer Pwr...

Page 15: ...APPENDIX B...

Page 16: ......

Page 17: ......

Page 18: ......

Page 19: ...APPENDIX C...

Page 20: ...ASB 4 ASB INTERIOR LABELS WARNING For continued protection against fire hazard replace with same type and rating of fuse only MODEL Z9000 Desktop 1P 2P Full Load 700 Watts 115 VAC 60 Hz Only Varitron...

Page 21: ...APPENDIX D...

Page 22: ...Specification Sheet Z9000...

Page 23: ...Specification Sheet Z9000 1P...

Page 24: ...Specification Sheet Z9001 2P...

Page 25: ...Specification Sheet FW9001 2P...

Page 26: ...Specification Sheet Z9001 2P...

Page 27: ...Specification Sheet Z9001 2P...

Page 28: ...APPENDIX E...

Page 29: ...PHOTOS OF Z9000 SERIES OPHTHALMIC CONTROL SYSTEMS...

Page 30: ...PHOTOS OF ACCESSORIES...

Page 31: ...PHOTOS OF ACCESSORIES...

Reviews: