TECHNICAL INFORMATION
and then screw the gas purge valve (with a torque
of 2.5 Nm) as shown in the following figure.
TV 301 CONTROLLER INSTALLATION
The controller can be mounted in two position:
−
bottom mounting (as per the complete system)
−
side mounting.
To install the controller execute the following pro-
cedures.
Bottom mounting
See the following figure.
1. Turn the pump upside-down;
2. Screw the 4 fixing studs (provided with the ac-
cessories bag) in the holes on the bottom of
pump;
3. Place the controller on the studs aligning the 4
rubber holes on the top of the controller case;
4. Press the controller towards the pump;
SERIAL CABLE INSTALLATION
The supplied serial cable must be installed when
the TV 301 Navigator or the optional vent valve
have to be controlled by means of a remote com-
puter.
5. Turn the pump again;
6. Plug the line cord and connect the pump cable
to the pump;
The cable is installed fixing the 9 pin D-type con-
nector into the P2 serial connector as shown in the
following figure.
7. Connect the mating connector with the jumper
on the interlock signal to start the pump.
MOUNTING 1
42
87-900-946-01(B)