TECHNICAL INFORMATION
33
87-900-924-01(B)
Example:
−
writing request of logic window n. 10 with the
logic value ‘0’ (OFF):
MASTER to SLAVE:
<0x02> + <0x80> +
<’010’> + <0x31> + <0> + <0x03> +
<’82’>
SLAVE to MASTER:
<0x02> + <0x80> +
<0x6> + <0x03> + <’B2’>
When the
MESSAGE
has a wrong
<CRC>
,
the
SLAVE will respond with the following string:
<STX>+<ADDR>+<NACK>+<ETX>+<CRC>
where
<STX>, <ADDR>, <ETX>
and
<CRC>
have the meaning previously discussed, while
<NACK>
= 0x15.
When the
MESSAGE
requires an access to an un-
known window, the SLAVE will respond with the
following string:
<STX>+<ADDR>+<’2’>+<ETX>+<CRC>
where
<STX>, <ADDR>, <ETX>
and
<CRC>
have the meaning previously discussed.
When the
MESSAGE
is a writing command to a
window, but the data is of a wrong type, the SLAVE
will respond with the following string:
<STX>+<ADDR>+<’3’>+<ETX>+<CRC>
where
<STX>, <ADDR>, <ETX>
and
<CRC>
have the meaning previously discussed.
When the
MESSAGE
is a writing command to a
window, but the data has a value out of the admit-
ted range, the SLAVE will respond with the follow-
ing string:
<STX>+<ADDR>+<’4’>+<ETX>+<CRC>
where
<STX>, <ADDR>, <ETX>
and
<CRC>
have the meaning previously discussed.
When the
MESSAGE
is a writing command to a
reading only window, or it is a reading command of
a writing only window, the SLAVE will respond with
the following string:
<STX>+<ADDR>+<’5’>+<ETX>+<CRC>
where
<STX>, <ADDR>, <ETX>
and
<CRC>
have the meaning previously discussed.
Window Meanings
N. Read/
Write
Data
Type
Description Admitted
Values
000 R/W L Start/Stop
(in remote
mode the
window is a
read only)
Start = 1
Stop = 0
(default = 1)
008 R/W L Remote
(de-
fault) or Se-
rial
configura-
tion
Remote = 1
Serial = 0
(default = 1)
100 R/W L Soft
Start
(write only in
Stop condi-
tion)
YES = 1
NO = 0
(default = 1 the
first time and 0
the following)
101
R/W
N
Set Point 1
type
0 = Frequency
1 = Current
2 = Time
(default = 0)
102
R/W
N
Set Point 1
threshold
(expressed
in Hz, mA or
s)
Hz: 150 to 1050
mA: 0 to 500
s: 0 to 4x10
9
(default = 1000)
103
R/W
N
Set Point 1
delay: time
between the
pump start
and the set
point check
(seconds)
0 to 4x10
9
(default = 0)
104
R/W
L
Set Point 1
output sig-
nal when
the meas-
ured value
is greater
than the
threshold
one
0 = output signal
high
1 = output signal
low
(default = 0)
105 R/W N Set
point
1
hysteresis
(in % of
threshold)
0 to 100
(default = 2)
108
R/W
N
Baud rate
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
109 W L Pump
life/
cycle time/
cycle num-
ber reset
To reset write ‘1’
110 R/W L Interlock
type
(default = 1)
Impulse = 0
Continuous = 1
120 R/W N Rotational
frequency
setting (Hz)
150 to 1050
(default = 1050)