Operation
ProStar 363 Fluorescence Detector
27
Wavelength Range Setup Screen
Wave
Length
Press the Wave Length key
Specify an excitation wavelength
Ex Em (200-850, 0) nm
250 350
The currently specified excitation wavelength is displayed. (The
default value is 250 nm.) The desired excitation wavelength
(input value) can be entered (range = 200 to 850 nm).
Press the Numeric key for the desired excitation wavelength,
press Enter
→
Fluorescence wavelength setup screen.
If the indicated excitation wavelength is acceptable, simply press
the
▼
key to go to the fluorescence wavelength screen, or press
the
▲
key to return to the monitor screen. As an alternative,
press the Escape key to return to the monitor screen without
changing the preset setting.
Specify a fluorescence wavelength
Ex Em (250-900, 0) nm
250 350
The currently specified fluorescence wavelength is displayed;
the default value is 350 nm. The desired fluorescence
wavelength (input value) can be entered (range = 250 to 900
nm).
Press, Numeric key for the desired excitation wavelength, the
Enter key
→
Monitor setup screen
If the indicated fluorescence wavelength is acceptable, simply
press the
▼
key to go to the monitor screen, or press the
▲
key