4
03-914875-00:1
You will receive the instrument and the accessory package in a
single shipping carton. Before you proceed with the unpacking,
visually check the outside of the carton for any evidence of
shipping damage, i.e., water stains, crushed corners, etc. If such
evidence is present do not unpack the contents but report the
condition to the carrier, and to:
Varian Chromatography Systems
2700 Mitchell Drive
Walnut Creek, California 94598-1675
Attention: Manager of Customer Service
Phone (925) 939-2400
or your local Varian Sales/Service Center
Unpacking
WA R N IN G
Avoid back strain or injury by following all precautions
for lifting heavy
Carefully unpack the instrument and place it on the bench top.
After unpacking the system, carefully check the contents of each
box against the packing list.
If any part is missing or damaged, contact the nearest Varian
dealer.
Laboratory Requirements
The instrument should be installed in a laboratory that meets the
following requirements:
Power Supply
Line voltage: 100 to 115 or 220 to 240 Vac
±
10% of the rated
voltage.
Frequency:
50 or 60 Hz
±
4% of the rated frequency.