Model 990 dCLD II Component Leak Detector
5-7
3. Remove the KF-25 (ISO NW-25) quick clamp
➀
to disconnect the spectrometer tube
from the Turbo Spec Tube Assembly (Figure 5-2).
Figure 5-2 KF-25 Quick Clamp
4. Lay the spectrometer tube down on a clean non-magnetic surface.
CAUTION
If the spectrometer tube magnet comes in contact with a
magnetic surface, the magnet may lose its gauss causing the
spectrometer tube to lose sensitivity.
CAUTION
Do not remove the 6 magnet body bracket screws
➀
shown in
Figure 5-3 while performing maintenance on the spectrometer
tube.
Figure 5-3 Magnet Body Bracket Screws
➀
➀