User Manual: UM50566-EN
Optica 30/40/50 series
Revision: 3.0
Date of release: 2019-01-09
30/36
www.vareximaging.com
6.
Technical specification
Materials
Stainless steel, aluminium, copper, lead, brass, leaded
bronze, plastics (ABS)
Power source
24 V DC ± 10%
Power consumption
35 VA (Optica 30/50 series only)
80 VA (Optica 40 series only)
X-ray rating up to
150 kVp
Application
Stationary equipment for radiography
Collimation
Square field multilayer
Mounting plane distance from focus
80 ± 1 mm
Aluminum equivalent contribution to
total filtering
(X-ray beam = 75 kV)
Min. Al 1.2 mm
Additional Filtration
(X-ray beam = 75 kV;
EN60601-1-3: §7.3, §7.5)
1 mm Al + 0.1 mm Cu
1 mm Al + 0.2 mm Cu
Combined 2 mm Al + 0.3 mm Cu
Manual selection and remote control (both motorized)
(
CF002 configuration only
)
Confinement of extra-focal radiation
(Set focus distance, SID 1 m;
EN 60601-2-54: §203.8.4)
< 150 mm
Square X-ray field selection
Min: 00 x 00 cm
Max: 48 x 48 cm (± 1 % at 1 m SID)
(to reach 43 x 43 cm at 90 cm SID)
Motor controlled shutters
Yes
Motor controlled additional filtration
CF002 configuration only
Knobs for manual setting
Yes
(
30/40 Series only
)
Input for additional setting
Remote control (CAN-bus)
Touch screen (
40 Series only
)
Light field source
White power LED
Timer and control to limit light source
exposure time to prevent overheating
Yes
(default 20 sec, adjustable)
Light field indicator illumination
(Illumination at 1m from the focus,
Field size at 35 x 35 cm;
EN 60601-2-54: §203.8.102.5)
≥ 230 lx, at 100 % power setting of light field source
Light field indicator-contrast
(Edge contrast setting 35 x 35 cm
at 1m SID ;
EN 60601-2-54: §203.8.102.5)
> 4 : 1
Light field indicator-Accuracy
(EN 60601-2-54: §203.8.102.6)
< 2 % SID
Linear light projection
(To align collimator with image receptor)
Laser Class 1 (IEC 60825-1) (optional feature)
Laser Class 2 (IEC 60825-1) (optional feature)