4343W/4336W-G5/2530W-G5 X-ray Detectors
Table 16
Problems and Solutions (Continued)
Acquired image does not respond
to exposure or the subject that is
being X-rayed does not show up in
the acquired images
1.
Let power-off (
Section 11.7
) or reboot the detector (
Section 11.8
),
re-insert the battery re-establish the link.
2.
Ensure that the tube dose is set to the correct settings and that
the detector is in the path of the X-ray beam.
3.
Acquire another image at the recommended median count as
stated in
Section 11.12.3
.
4.
Ensure that the X-ray detector is in the X-ray Beam path.
5.
If problem persists, consult the manufacturer or field service
technician for help.
Acquired image shows vertical or
horizontal lines
1.
Let power-off (
Section 11.7
) or Reboot the detector (
Section
11.8
), re-insert the battery, re-establish the link.
2.
Ensure that the tube dose is set to the correct settings.
3.
Acquire another image at the recommended median count as
stated in
Section 11.12.3
.
4.
If the problem persists, add the defects to the defect map and
redo a gain calibration.
5.
Acquire another image.
6.
Move the X-ray detector away from any device that may
introduce noise.
7.
If problem persists, consult the manufacturer or field service
technician for help.
Residual X-ray image from previ-
ous acquired image shows in cur-
rent acquired image
Charge on the sensor pixels from an exposure beyond the
saturation point of the pixels may cause a residual image. It can be
erased by increasing time between exposures or taking multiple
images until the artifact fades away.
Out of virtual memory
Close some of the windows that are currently open.