3101126 Rev A Vapotherm Oxygen Assist Module Instructions for Use
Page 15
•
Do not sterilize the sensor or patient cables by irradiation, steam, or ethylene oxide. Consult the
cleaning and sterilization instructions in the documentation for reusable pulse oximetry sensors
and patient cables.
Note:
The sensors used with this device can be categorized as surface devices contacting
skin for a limited period of time. The sensors have passed the recommended biocompatibility
testing and are therefore in compliance with ISO 10993–1.
SpO
2
Parameter Box Display
Note: The indicator lights provide information on Signal IQ. Green denotes Signal IQ values of equal to
or greater than 0.3%. Red denotes Signal IQ values of less than 0.3%.
The SpO
2
display range is from 1-100%. A value of “---“ indicates that no measurement is available. The
pulse rate display range is from 1-240 bpm. A value of “---“ indicates that no measurement is available.
Note:
The SpO
2
, pulse rate, and Signal IQ are updated once per second.
SpO
2
Target Setup
1.
Press the SpO
2
Target Box on the Home screen or select the Patient menu to set the SpO
2
target.
2.
Press the %SpO
2
Target parameter box on the touch screen, and set the patient’s target SpO
2
using the settings control knob.
3.
Set the Target Range by pressing the %SpO
2
Upper and Lower Range parameter boxes on the
touch screen, and set the percent difference from the %SpO
2
Target using the settings control knob.
NOTE: Setting the Target Range does not change the SpO
2
Target. This setting only adjusts
the time in range calculation when reviewing trends.
Section 9 SpO
2
Setup