Page 10
3101126 Rev A Vapotherm Oxygen Assist Module Instructions for Use
Section 4 Assembly and Connections
System Assembly
1.
Affix the Oxygen Assist Module to the Pole Mount Assembly.
2.
Mount the Oxygen Assist Module to the Vapotherm approved roll stand as shown.
3.
Remove the silicone plug in the oxygen sensor cover on the back of the Precision Flow.
4.
Plug the communication cable into the Oxygen Assist Module and tighten the two thumb screws.
5.
Connect the communication cable from the Oxygen Assist Module into the Precision Flow’s
interface port.
6.
Plug the Oxygen Assist Module Power Supply into the Oxygen Assist Module and into an
electrical outlet marked “Hospital Grade” or “Hospital Only.”
NOTE: Ensure proper orientation on Power Supply connector when inserting into the Oxygen
Assist Module.
WARNING:
Do not position the Oxygen Assist Module so that it is difficult to disconnect the
device.
Sensor Connection
•
Connect the SpO
2
patient cable to the SpO
2
receptacle on the connector panel of the Oxygen
Assist Module. Please consult Attaching the Sensor to the Patient Cable in the Masimo
DIRECTIONS FOR USE
to learn how to connect the sensor.
Oxygen
Assist
Module
mounted
and
plugged
into the
Precision
Flow and
Power
Vapotherm Oxygen Assist Module Connector Panel