A-3
2 Guidance and Manufacturer’s Declaration – Electromagnetic
Immunity
The MEDIC Vaporizer is intended for use in the electromagnetic
environment specified below. The customer or the user of the
MEDIC Vaporizer should assure that it is used in such an environ-
ment.
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Immunity
test
± 8 kV contact
discharge
± 15 kV air
discharge
± 8 kV contact
discharge
± 15 kV air
discharge
IEC 60601
test level
Compliance
level
Electromagnetic
environment –
guidance
Floors should be
wood, concrete or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines
± 2 kV for
power supply
lines
± 1 kV for
input/output
lines is not
applicable
Mains power quality
should be that of a
typical residential,
commercial or
hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Mains power quality
should be that of a
typical residential,
commercial or
hospital
environment.
INFORMATION ON ELECTROMAGNETIC COMPATIBILITY
A
pp
end
ix