24
6 C L E A N I N G / H YG I E N E
6.1 Mouthpiece
X
Medical practice and clinic:
Before the first use and after each use,
prepare Mouthpiece for uncritical or semi-critical A - products (DGKH)
in compliance with the generally recognized procedures of hospital hy-
giene.
X
Domestic use:
Clean Mouthpiece before the first use as well as after each
use for at least five minutes in a lukewarm (40 °C - 50 °C ( 104 °F - 122 °F))
dishwashing detergent bath (use standard household detergent).
6.2 Cooling Unit
The Cooling Unit with Mouthpiece is intended for one person and can be used
by that person several times. They must be free of deposits. Deposits on the
inside of the Cooling Unit are caused by steam condensate. Cleanliness is a
prerequisite for proper function.
X
Check the Cooling Unit with Mouthpiece prior to each use.
X
Clean the Cooling Unit with Mouthpiece no later than 7 days after the last
use and if deposits are visible on the inside of the Cooling Unit.
6.2.1 Disassembling the Cooling Unit
X
Heat up the Cooling Unit for better disassembly, as the steam conden-
sate can be sticky.
X
Pull the Cap Lock towards the
Mouthpiece.
X
Pull the Cooling Unit Cap off the Va-
porizer.