SPECIFICATIONS
Latest amendment date: 21/10/2011
CE 0120
Model Number
PY-001
Power Supply
1.5VX 2, AA Size Alkaline
battery
Power Consumption
< 2W
Active current
consumption
<650 mA
Standby current consumption
<0.35mA
Battery life
Continue operation:
1 hour (min)for brand new 1.5V X
2,AA, Alkaline battery or Ni-MH
rechargeable Batteries (option).
Full battery indication
>2.2V, green light
Weak battery indication
2.0V ~2.2V, orange light
Low battery indication
<2.0V, red light
Operation environment
10~40°C, 30%~80% RH
Storage/ Transportation
environment
-20~70°C,
≦
85% RH
Nebulization rate
0.2mL/min,~1.0mL/min(Max)
Particle size
<5um
Time AUTO-OFF
Power on, operate 20 min. Then
auto-off
Nebulizer auto-off
No medication
Medication bottle volume
2~8 mL (Max.)
Ultrasonic Oscillation
Frequency
103 KHZ~123 KHZ ± 1K
Reusable parts life
Medical cup, mask,
mouthpiece:
one year
According as reliability test
result-
One time per day
The specifications may be changed without prior notice.
* The product has certified to meet the Harmonized Standard
EN60601-1(Product
Safety) and EN60601-1-2 (Electromagnetic
Compatibility)
* The grounding mechanism