MB0500400/ MB0500500

Instruction Manual

Portable Nebulizer

POCKET AIR

R

Summary of Contents for MB0500400

Page 1: ...MB0500400 MB0500500 Instruction Manual Portable Nebulizer POCKET AIR R ...

Page 2: ... of Symbols 5 Package Contents and Overview 6 How to Operate with Batteries for Model MB0500500 8 How to Operate with Adapter for Model MB0500400 10 Cleaning and Disinfection 12 Troubleshooting 14 Specifications 15 Technical Data 16 Optional Parts and Accessories 21 Table of Contents ...

Page 3: ...y AA NiMH rechargeable batteries optional item not included instead of disposable batteries This device has extra power option it can be driven by both two AA batteries and power adapter optional item not included With additional power adapter helps you to save battery costs in long term Our devices are not only brings you conveniences of lightweight compact and portable the virtually silent opera...

Page 4: ...s device and the content of instruction manual Intended patients This device design for all ages of patients Except for the patients who are unconscious not breathing spontaneously or having pulmonary edema Recommended operation environment This device is intended for using in the medical facility such as hospital clinic and doctor s office a room of general household and open air environment with...

Page 5: ...t there is a risk of spreading infectious illness Clean all parts of your nebulizer before use after each use and after extended storage Adult supervision is required when this device is used by children and individuals who require special assistance Never operate this device if any of the parts are not working properly or have been damaged Be sure the device has been properly cleaned before use t...

Page 6: ...n DC Power source Direct Current Operating temperature limits 10 C to 40 C Storage and delivery temperature limits 20 C to 70 C Operating humidity limits 30 to 85 R H Storage and delivery humidity limits 20 to 75 R H Manufacturer Date of manufacture Serial number Batch code Disposal of Electrical Electronic Equipment WEEE Do not treat this product as household waste This device compiles with the r...

Page 7: ...aged parts before you use this device In the case of missing parts malfunction or damage please contact the store where you purchased from or the nearest dealer 1 Main unit MB0500400 MB0500500 MB0500301 4 Mask Single use Do NOT recycle 2 Medication cup 3 Instruction Manual MB0500400 MB0500500 Instruction Manual Portable Nebulizer POCKET AIR R ...

Page 8: ...tor Battery Cover Main Unit Power Cable Jack Power Cable Jack Cover Bottom View Model no MB0500400 Bottom View Model no MB0500500 Light Indications Press ON OFF Button to turn on the power Press Release Button to remove the medication cup Green light on Ongoing operation Orange light on flash Low power Malfunction No light Malfunction ...

Page 9: ... and pull up the battery cover to open it Insert two AA alkaline NiMH rechargeable batteries according to the battery polarity indicated inside the battery compartment Put down and push forward the battery cover to lock it Open the medication cup cap Fill the medication cup with saline DO NOT EXCEED THE LIMIT Max 6ml Close the medication cup cap properly Align the three rubs on the back of the med...

Page 10: ...d the mouthpiece If you use a mask place the mask over your mouth and nose Breathe in and out slowly through your mouth until aerosol formation stop Press ON OFF button to turn it off Remove the batteries from the main unit ON OFF Button Battery life and battery replacement 1 Battery life is usually depending on the capacity and condition of the batteries In general this device can operate about 4...

Page 11: ...vices it is recommended that keep the product unplugged when not in use 1 Set up the medication cup Align the three rubs on the back of the medication cup with three trenches of the main unit to insert the medication cup into the main unit 2 Fill the saline 3 Connect with adapter Open the medication cup cap Fill the medication cup with saline DO NOT EXCEED THE LIMIT Max 6ml Close the medication cu...

Page 12: ... the mouthpiece between your teeth with your lips firmly sealed around the mouthpiece If you use a mask place the mask over your mouth and nose Breathe in and out slowly through your mouth until aerosol formation stop Press ON OFF button to turn it off Remove the adapter from main unit and electrical outlet ON OFF Button 5f5P CC ...

Page 13: ...n unit 5 Remove the medication cup from the main unit 6 Wash and rinse the medication cup with distilled water 7 Shake off excess water and allow parts to be fully air dried on a clean dry towel 8 Use gauze or clean towel to wipe off stains on the main unit if necessary 9 Make sure that all cleaned parts are completely dry before you store them or use them next time 1 Keep the battery compartment ...

Page 14: ...ry leakage 3 Do NOT leave the device under direct sunlight in high humidity extreme heat or cold environment 4 Keep this device away from fire high electromagnetic fields and out of the reach of children Carrying and storing Medication cup Main unit Power adapter only available on Model MB0500400 The adapter figure might be different from above The following items are able to be disinfected by boi...

Page 15: ...e adapter to main unit No atomization even if power switch and indicator are on The medication cup is not installed properly Refer to Set up the medication cup procedure to re install the medication cup The mesh holes are clogged Refer to Cleaning and Disinfection procedure to clean the medication cup The mesh is broken Replace the medication cup Power indicator turns into orange light constantly ...

Page 16: ...ct Portable Nebulizer Model MB0500400 MB0500500 Method of operation Active Vibrating Mesh Technology Power supply AA Alkaline Battery x 2 AA NiMH rechargeable Battery x 2 Power Adaptor Only Available on mode MB0500400 Power consumption 1 5W Vibrating frequency Approx 130 kHz 15 Nebulization rate 0 25 ml min Particle size MMAD 5um Capacity of medication cup Approx 6 ml Dimension Approx L76 5 X W41 ...

Page 17: ...nnex CC 3 using the multistage cascade impactor method to measure the particle size Particle Size MMAD 4 21um MMAD Mass Median Aerodynamic Diameter Medication Cup Capacity 6 ml max Noise Less than 50dB Around 1 meter 0 11 10 100 AeroDynamic Diameter µm 0 20 40 60 80 100 Cumulative Fraction 0 10 20 30 40 50 Mean Mass Fraction individual tests mean ...

Page 18: ...the electromagnetic environment specified below The customer or the user of the MB0500400 MB0500500 should assure that it is used in such an environment Emission test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The MB0500400 MB0500500 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference ...

Page 19: ...to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in...

Page 20: ...xed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE1 At 80 MHz and 800 MHz the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and refl...

Page 21: ...the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where p is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE2 These guidelines may not apply in all situations Electr...

Page 22: ...tion cup and masks are capable of Model No MB0500400 MB0500500 Medication Cup MB0500301 Mask Child MBPN00204 Mask Adult MBPN00203 Power Adapter MBPN00205 MicroBase Technology Corp No 756 Jiadong Rd Bade Dist Taoyuan City 33464 Taiwan MedNet GmbH Borkstrasse 10 48163 Muenster Germany REP EC MB0500400_MB0500500_T11 16006 01 ...

Page 23: ...ain unit of this Portable Nebulizer 3 During the warranty period we will assume the expenses arising from maintenance work for the defects and malfunction of the main unit of this Portable Nebulizer due to normal wear and tear excluding any breakdown resulting from improper use Any expenses of non maintenance work including but not limited to the back and forth shipping cost or expenses arising fr...