background image

 

 

 

NOTE:

 

If amber  charge status  light  fails to  illuminate after connection, check to  see that the 
battery pack is properly connected to the charger.  If green power “ON” indicator light 
flashes, leave battery pack on charger for one hour.  If amber charge status light still 
fails to illuminate after one hour, disconnect battery pack from charger and contact an 
authorized Vante® SEBRA service center for battery pack replacement. 

3.3  Leave battery pack on battery charger until completely charged.  Battery pack is completely 

charged when amber charge status indicator light flashes every ¼ second. 

Charge status is represented by the amber charge status indicator light as follows: 

Charge Status  Charge Indicator Light Activity  Description 

Battery Absent  Amber light not illuminated 

Battery pack not connected 

Pending Mode  Amber light flashes every 1½ 

second 

Charger in a slow charging 
mode.  Maintain battery pack 
on charger.  Charger will 
automatically switch to Fast 
Charge when appropriate.   

Fast Charge 

Amber light continuously 
illuminated 

Charger in fast charge mode 

Float Charge 

Amber light flashes every ¼ 
second 

Fast charging complete, 
battery pack ready for use 

 

 

 

CAUTION:

  If    Fast  Charge  period  (amber  charge  status  light  is  continuously  illuminated) 

exceeds  6  hours,  disconnect  battery  pack  from  battery  charger  and  contact  an 
authorized Vante® SEBRA service center. 

 

If  Pending  Mode  period  (amber  charge  status  light  flashes  every  1½  second) 
exceeds one hour, disconnect battery pack from charger and contact an authorized 
Vante® SEBRA service center for battery pack replacement.

 

3.4 

Disconnect the battery pack from the charger and reinstall into the RF power source or set 
aside as a back-up.  Disconnect battery charger from AC power source (wall outlet).  

NOTE:

  It  is  important  to  remove  battery  pack  from  the  charger  before  disconnecting  the 

charger from the AC power source, or battery pack will lose its charge. 

 

Battery  packs  not  in  active  use  should  be  maintained  by  discharging  and  recharging 
every 30 days. 

 

 

Summary of Contents for SEBRA 2389

Page 1: ...Model 2389 RAPID CHARGER For use with the Model 2380 Mini Hand Held Tube Sealer INSTRUCTION MANUAL ...

Page 2: ......

Page 3: ...i 1997 SEBRA Vante Inc 3480 E Britannia Dr Suite 120 Tucson AZ 85706 USA Telephone 520 881 6555 Fax 520 323 9055 Toll Free 877 565 5557 Part No VT 23892000 01 Rev B ...

Page 4: ...lated procedures are used for purposes other than those stipulated herein validation of the specific application should be obtained otherwise Vante SEBRA assumes no liability or obligation nor guarantees product performance The contents of this manual are not to be construed as license to operate under nor a recommendation to infringe upon any patents Personnel using the product do so at their own...

Page 5: ...mponent Identification 1 2 0 Battery Pack Removal from Power Source 2 3 0 Battery Charging 3 4 0 Battery Pack Insertion into Power Source 5 Table of Figures Figure 1 1 Components of Rapid Charger 1 Figure 2 1 Accessing and Removing the Battery Pack 2 Figure 4 1 Opening Battery Compartment Door 5 Figure 4 2 Battery Connector Location 5 Figure 4 3 Battery Installation 6 Figure 4 4 Battery to Power S...

Page 6: ......

Page 7: ...arger for use with battery packs of the Model 2380 Mini Hand Held Tube Sealer Model 2380 The Rapid Charger comes with a power cord 1 2 Component Identification Shown in Error Reference source not found are the component parts of the Rapid Charger Model 23891000 01 Charger Assembly Power Cord Model number depends upon country of use Figure 1 1 ...

Page 8: ...ead and or tubing are dry and clean to eliminate the possibility of further arcing Turn the power source on and continue sealing If the battery low condition is indicated again and arcing has not occurred then it is time to charge the battery pack 2 2 Prior to battery pack removal turn the power switch to the off position Disconnect the sealing head and RF power cable and remove the power source f...

Page 9: ...ack NOTE Check to see that the power cord plug matches the power receptacle for the country in which the battery charger is being used If it does not or if the green power ON indicator light does not illuminate contact an authorized Vante SEBRA service center 3 2 Once the battery pack is removed from the RF power source Section 2 0 plug the battery pack s red and black connector into the matching ...

Page 10: ...ry pack on charger Charger will automatically switch to Fast Charge when appropriate Fast Charge Amber light continuously illuminated Charger in fast charge mode Float Charge Amber light flashes every second Fast charging complete battery pack ready for use CAUTION If Fast Charge period amber charge status light is continuously illuminated exceeds 6 hours disconnect battery pack from battery charg...

Page 11: ...en the battery compartment door on the power source case by sliding the latch to the right See Figure 4 1 Figure 4 1 4 2 Position the battery connector as shown in Figure 4 2 CAUTION Use only with battery packs equipped with a Positive Thermal Cutout PTC device and or approved for use with Vante SEBRA Model 2380 Figure 4 2 ...

Page 12: ...rs are keyed to engage properly Be sure that the RED battery pack connector is engaged into the RED power source connector and the BLACK into the BLACK If they do not engage easily do not attempt to force the connectors together Improper battery connection will result in damage to the RF power source and will void the warranty Figure 4 4 4 5 Close and latch the battery compartment door If desired ...