8. Since the hole for the pipet tips is so large, the solution will gradually lose heat over time.  Testing has shown 
an approximately 4 degree reduction over 3 hours with 60

o

C water.  When the Controller was set at 190

o

C the 

sides of Circle Heater got to 78

o

C, so be careful when handling the Circle Heater.  

  

 

  

 

Figure 8. Pouring pre-heated solution in SpinVessel in the Circle Heater. Replacing the lid in the  

proper orientation for use with 96 channel pipet head.  Independent check of solution temperature. 

 
OPERATION of the SpinVessel  

Please  refer  to  the  technical  information  provided  in  Technote  378B. Please  contact V&P  Scientific  for  more 
information. 
 

TEMPERATURE CONTROLLER OPERATING INSTRUCTIONS

 

Supplied as an attached document. 

 

Summary of Contents for SPINVESSEL VP 418SV2-H-1LFB-CC

Page 1: ...igure 1 Basic Parts of Circle Heater SpinVessel VP 418SV2 H 1LFB CC SpinVessel Controller not shown WARNING The VP 418SV2 H 1LFB CC Circle Heater SpinVessel will be very hot when operating so exercise caution Do not handle red aluminum heating portion of the unit while it is operating it will be HOT Temperature Controller Circle Heater SpinVessel Motor Unit ...

Page 2: ...ter to Temperature Controller Load Power Sensor Load Sensor On Off Switch Temperature Controller SpinVessel Controller Temperature Sensor Temperature Controller Power Cord Temperature Controller Cable USB Drive USB Cable Temperature Sensor Hole Plugs SpinVessel Controller Power Supply and Cable SpinVessel Controller Power Cord SpinVessel Controller Cable Circle Heater SpinVessel SpinVessel SpinBas...

Page 3: ...n the free end for connecting to the Temperature Controller Figure 3 page 2 Connect the round 3 pin plug of the Connector Cable into the receptacle marked Load on the back of the Temperature Controller as shown in Figure 3 Also on the back of the Temperature Controller connect the 2 prong plug of the Controller Cable into the receptacle marked Sensor 3 See Technote 378B for instructions on setting...

Page 4: ...ust be careful to keep fingers away from the back of the Controller when doing this as the magnets are strong and fingers can be pinched 6 There are 2 holes for Temperature Sensor placement use the one that works best for the pipetting head to be used The other one can be covered with the provided plug Figure 5 Attach the external Temperature Sensor Controller Figure 6 Position the external Temper...

Page 5: ...Vessel and the heating element heat transfer must occur through the air and to a limited extent through the Spin Base of the SpinVessel motor unit Therefore the temperature setting for the Controller will be much higher than the actual temperature in the solution in the SpinVessel 5 It is best to pre heat the Circle Heater and the solution to be mixed as it takes a long time for the solution to co...

Page 6: ...ater got to 78o C so be careful when handling the Circle Heater Figure 8 Pouring pre heated solution in SpinVessel in the Circle Heater Replacing the lid in the proper orientation for use with 96 channel pipet head Independent check of solution temperature OPERATION of the SpinVessel Please refer to the technical information provided in Technote 378B Please contact V P Scientific for more informat...

Page 7: ... be free from defects in material and workmanship when used under normal laboratory conditions for one year This warranty begins on the date of delivery from V P Scientific In the event this product fails under normal laboratory conditions within the specified period of time because of a defect in material or workmanship V P Scientific will at its option repair or replace the product Damage to the...

Reviews: