Use Loctite 401 to secure the new grommet to the cover.
P3 – Faulty Main PCB Air Switches
IMPORTANT
NOTE:
There are no serviceable parts on the Main PCB. These air switches are not
serviceable and the whole board must be replaced if they fail.
As this is part of the pneumatic
system, it is important that all connections are airtight.
1.
If there are no leaks in the hand control, airline tubing and grommets, there could be a leak
in air tubes to the pressure switch or failure of air switch. Check by performing the air leak
test with the hand control buttons.
2.
If there is an air leak,
tighten all air tubes and recheck.
Air Switches
P-600 Technical Manual
REV: 101316
Page 14