17
Troubleshooting
For information about parts, accessories, or service -- contact your dealer directly or contact VanGuard
Microscopes at 1-800-423-8842.
I can’t see anything through the eyepieces/camera
Step #
Possible Cause
How to Check
Result
Solution
For More Info
1
Light not turned on
Visually inspect
Turned on
Go to next step
Not turned on
Turn on light
Page 10
2
VLC set too low
Visually inspect
Not set too low
Go to next step
Set too low
Increase VLC
Page 10
3
Objectives not installed
Visually inspect
Installed
Go to next step
Not installed
Install objectives
Page 7
4
Condenser not installed
Visually inspect
Condenser installed
Go to next step
Condenser not installed
Install condenser
Page 7
5
Prism slider pulled out/pushed
in (trinocular models only)
Visually inspect
Prism slider in correct
position
Go to next step
Prism slider not in
correct position
Pull out for camera/push in for eyepieces Page 12
6
Something blocking light path
between substage illuminator
and objectives
Visually inspect
Blockage
Remove blockage
No blockage
Contact dealer or VanGuard Microscopes Page 17
I can’t focus on the specimen
Step #
Possible Cause
How to Check
Result
Solution
For More Info
1
Up-stop set too low
Visually inspect
Set too low
Reset up-stop to appropriate height
Page 11
Not set too low
Go to next step
2
Using the 100X objective
without immersion oil
Visually inspect
Using oil
Go to next step
Not using oil
Use immersion oil
Page 13
3
Condenser not installed
Visually inspect
Condenser installed
Go to next step
Condenser not installed
Install condenser
Page 7
4
Oil in objective/dirty objectives
Visually inspect
Objectives clean
Go to next step
Objectives dirty
Clean or replace objectives
Page 15
5
Slide placed on stage upside
down
Visually inspect
Slide upside down
Flip slide
Slide not upside down
Go to next step
6
Cover glass too thick
Visually inspect
Cover glass too thick
Use #1.5 (0.17mm) cover glass
Cover glass not too thick
Contact dealer or VanGuard Microscopes Page 17
The stage keeps drifting down
Step #
Possible Cause
How to Check
Result
Solution
For More Info
1
Focus tension is set too low
Check focus tension
control knob (page 10)
Not enough tension
Increase tension
Page 10
Plenty of tension
Contact dealer or VanGuard Microscopes Page 17