The next three pages are dedicated to assembling a working microscope.  

The following section “Using your 1200CM Series Microscope” (starting on 

page 10) explains the various features of the microscope and how to use 

them.

After removing the microscope parts from the protective foam and plastic 

packaging and checking it over for all components and accessories (see 

list on page 3), you can begin assembly.
Place the stand on a stable counter top.

NOTE:  Do not release the head until it is firmly secured with the 

head set screw.

Loosen the head set screw (circled in figure 1) and rotate the head 180° 

so the eyepieces face forward.  
Retighten the head set screw.

Assembly

Setup

1

7

6

5

4

3

7

Figure 3

Figure 2

Figure 1

Remove the dust caps from the eyetubes, then insert  the eyepieces into 

the eyetubes (see figure 2).

After removing the objectives from their storage containers, individually 

install each one into the nosepiece by threading them clockwise into the 

threaded holes of the nosepiece (see figure 3).  It is generally considered 

best practice to install the objectives in numerical order.

The next four steps describe how to install the condenser.  With phase-

contrast  models  the  condenser  comes  pre-installed.    If  this  is  the  case, 

assembly is complete.  Skip ahead to “Centering the Condenser” on page 

8 unless you need to reattach your condenser for any reason.

Note:  The phase contrast condenser is pictured, but the assembly 

instructions are identical for the brightfield condenser.

Raise the substage and stage to their maximum height (see figure 4).  

Raise the stage via the coarse focus control and the substage using the 

substage adjustment knob.

Loosen  the  condenser  set  screw  enough  to  allow  the  neck  of  the 

condenser to slide through the silver ring.

8

All Models

Figure 4

Summary of Contents for 1200CM Series

Page 1: ...Covering Models 1210CM 1211CM 1220CM 1221CM 1222CM 1223CM 1230CM 1231CM 1232CM 1233CM Operation Manual 1200CM Series...

Page 2: ...25 N A Abbe condenser Phase contrast models come with a 1 25 N A Zernike condenser with phase annulus rings for 10X 25X 40X and 100X also has a 0 setting for brightfield work All condensers are mount...

Page 3: ...0X Plan Phase Objective 1 ea 20X Plan Phase Objective 1 ea 40X Plan Phase Objective 1 ea 100X Plan Phase Objective 1 ea Dispersion Filter 1 ea Blue Filter 1 ea Green Filter 1 ea Model 1230CM Trinocula...

Page 4: ...xels for 1210 and 1220 series 1200 CDPC 3 USB DIgital Camera System 3 Megapixels for 1230 series Other Accessories Part Number Description Call for Part Numbers Eyepieces 10X 16X 20X Call for Part Num...

Page 5: ...tering Knobs Brightfield only Aperature Iris Diaphragm Adjustment Lever Fine Focus Control Substage Collector Base Stand Mechanical Stage Slide Holder Slide Holder Knobs Focusing Tension Control Knob...

Page 6: ...Fuse Holder Parts Accessories Phase Annuli Centering Knobs Phase Contrast Condenser Iris Diaphragm Adjustment Light Bulb Centering Mechanism Substage Collector Set Screw 6 Light Bulb Access Door Lock...

Page 7: ...eyetubes see figure 2 After removing the objectives from their storage containers individually install each one into the nosepiece by threading them clockwise into the threaded holes of the nosepiece...

Page 8: ...er will reduce the field so that a small white hexagon is visible within a black field see Figure 6A Focusing of the hexagon is performed by turning the coarse fine focus controls This white hexagon i...

Page 9: ...nser assembly until the two rings of light are centered upon one another see Figure 8 Once the phase rings are centered remove the phase contrast centering telescope and replace with the eyepiece The...

Page 10: ...to induce contrast into a specimen not to adjust light intensity The mechanical stage X Y controls located underneath the right hand side of the stage see figure 9 provide easy and accurate positionin...

Page 11: ...icroscope Interpupillary and Diopter Adjustment 1 2 3 Interpupillary adjustment the distance between eyepieces is made through a folding action The Seidentopf design allows for a folding adjustment wh...

Page 12: ...a adapter assembly until it is firmly secured with the camera port set screw Tighten the camera port set screw Pull the prism slider completely out to divert the image to the camera port The prism sli...

Page 13: ...Models The 100X objective which comes with this microscope must be used with immersion oil in order to maintain image quality After use the objective tip needs to be wiped clean so that no oil residu...

Page 14: ...lose the light bulb access door and screw the knob back in to lock Return the microscope to normal operating position If necessary recenter the illuminator following the instructions on page 9 1 2 7 6...

Page 15: ...r ordering information The same care instructions apply to all optical parts on this VanGuard Microscope including the substage collector and the condenser lenses All other parts can be cleaned with a...

Page 16: ...can t get enough light Step Possible Cause How to Check Result Solution For More Info 1 Variable lighting control VLC is turned down too far Visually inspect VLC turned up Go to next step VLC turned...

Page 17: ...blockage Contact dealer or VanGuard Microscopes Page 17 I can t focus on the specimen Step Possible Cause How to Check Result Solution For More Info 1 Up stop set too low Visually inspect Set too low...

Page 18: ...25 NA 8 0mm WD 1 8mm FOV Achromatic or Plan Achromatic 20X 0 40 NA 0 50mm WD 0 9mm FOV 40X 0 65 NA 0 40mm WD 0 45mm FOV 100X oil 1 25 NA 0 25mm WD 0 18mm FOV Stage Dimensions 125mm x 135mm Stage Motio...

Page 19: ...19 Notes...

Page 20: ...Printed in the USA No VGMNL070319 1200CM...

Reviews: