RV16 for 1.666 volts at pin 11 of the
IOX connector.
15) Apply 15.00 volts and 1.0000 amps
and adjust RV11 for .1666 volts at pin
11 of the IOX connector.
16)
Repeat steps 14 and 15 for minimum
interaction.
8-6-2. Channel B IOX Calibration
1) Connect the 2300 as shown in Figure 8-8.
Select
φ
B, 150 (15) volts, and 1 amp
ranges.
2) Apply 150.00 volts and adjust RV17 for
5.000 volts at pin 4 of the IOX connector.
3) Apply 15.00 volts and adjust RV3 for
.5000 volts at pin 4 of the IOX connector.
4) Repeat steps 2 and 3 for minimal
interaction.
5) Apply 1.0000 amps and adjust RV18 for
5.000 volts at pin 5 of the IOX connector.
6) Apply .1000 amps and adjust RV4 for
.5000 volts at pin 5 of the IOX connector.
7) Repeat steps 5 and 6 for minimal
interaction.
8) Apply 150.00 volts and 1.0000 amps and
adjust RV19 for 5.000 volts at pin 6 of the
IOX connector.
9) Apply 15.00 volts and 1.0000 amps and
adjust RV8 for .5000 volts at pin 6 of the
IOX connector.
10) Repeat steps 8 and 9 for minimal
interaction.
11) Apply 150.00 volts and 1.0000 amps
and adjust RV20 for 1.666 volts at pin
11 of the IOX connector.
8-6-3. Channel C IOX Calibration
1) Connect the 2300 as shown in Figure 8-9.
Select
φ
C, 150 (15) volt and 1 amp
ranges.
2) Apply 150.00 volts and adjust RV21 for
5.000 volts at pin 7 of the IOX connector.
3) Apply 15.00 volts and adjust RV5 for
.5000 volts at pin 7 of the IOX connector.
4) Repeat steps 2 and 3 for minimal
interaction.
5) Apply 1.000 amps and adjust RV22 for
5.000 volts at pin 8 of the IOX connector.
6) Apply .1000 amps and adjust RV6 for
.5000 volts at pin 8 of the IOX connector.
7) Repeat steps 5 and 6 for minimal
interaction.
8) Apply 150.00 volts and 1.0000 amps;
adjust RV23 for 5.000 volts at pin 9 of the
IOX connector.
9) Apply 15.00 volts and 1.0000 amps and
adjust RV9 for .5000 volts at pin 9 of the
IOX connector.
10) Repeat steps 8 and 9 for minimal
interaction.
11) Apply 150.00 volts and 1.0000 amps
and adjust RV24 for 2.500 volts at pin
10 of the IOX connector.
12) Apply 150.00 volts and 1.0000 amps
and adjust RV25 for 1.666 volts at pin
11 of the IOX connector.
8-7.
Option IO-3 Calibration
Option IO-3 calibration should be performed
only after the 2301 has been completely
calibrated. The IO-3 may be calibrated in any
volts and any current range desired, provided
that the user provides full scale voltage and
current for those ranges (i.e. 600V on the
600V range). In the following procedure the
150V/1A ranges and inputs are used.
Connect the inputs as shown in Figure 8-7 and
select the 150V and 1A ranges. A DVM with
at least .03% DC accuracy is used to
monitor the outputs of Option IO-3.
The pin functions for the IO-3
Summary of Contents for 2300
Page 3: ...This manual covers the following Valhalla Scientific products Models 2300 2301 2300L and 2301L...
Page 19: ......
Page 20: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 50: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 74: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 100: ......
Page 101: ......
Page 103: ......
Page 106: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 113: ......
Page 114: ......
Page 115: ......