6-4-23.
Three-phase Four-Wire PT
Source Power Connections
Some applications require measuring the
power from a source rather than the power
into a load. These connections are shown in
Figure 6-23. When the Model 2300 is used in
this manner, the watts reading should be
multiplied by minus one (-1) and the PT ratio.
Observe the polarities of the potential
transformers.
Caution!
Do not exceed the
common mode specifications of the Model
2300.
6-4-24.
Three-phase, Four-Wire CT-PT
Source Power Connections
Some applications require measuring the
power from a source rather than the power
into a load. These connections are shown in
Figure 6-24. When the Model 2300 is used in
this manner, the watts reading should be
multiplied by minus one (-1) and the CT-PT
ratios. Observe the polarity of the current and
potential transformers. For best performance
the CURRENT COMMON terminals should
be grounded.
6-5. Single-Phase
Measurements
The power consumed by a single-phase load
or the power supplied by a single-phase
source may be easily measured with a Model
2300 or 2301. Three independent single-
phase measurements can be made at the same
time with the 2300. The following guidelines
should be used when making a single-phase
power measurement:
1) Connect the 2300 as shown in Figures 6-1
to 6-8 as determined by your application.
Make connections to the appropriate
current shunt terminal for the expected
range of current levels.
2) Press the appropriate
φ
A,
φ
B, or
φ
C
display selection push-button for the
channel connected.
3) Select the lowest current range that
provides the required display resolution
without the current peak overload LED
illuminating.
4) Select the lowest voltage range that
provides the required display resolution
without the voltage peak overload LED
illuminating.
5) Read the voltage, current and power on
the three displays. When measuring the
power supplied by a source, multiply the
power measurement by negative one (-1)
to obtain the correct watts polarity.
6) The displays may be frozen indefinitely
by pressing the HOLD push-button.
6-6. Three-Phase
Three-Wire
Measurements
The power consumed by a three-phase three-
wire connected load or the power supplied by
a three-phase three-wire source may easily be
measured with a 2300. The following
guidelines should be used when making these
measurements (also refer to AN104):
1) Connect the 2300 as shown in Figures 6-9
to 6-16 as determined by your application.
Make connections to the appropriate
current shunt terminals for the expected
range of current levels.
2) Press the 3
φ
3-WIRE display selection
push-button.
3) Select the lowest current range without
illuminating any of the peak current
overload indicators.
4) Select the lowest voltage range without
illuminating any of the peak voltage
indicators.
5) The total true power may be read
on the watts display.
Summary of Contents for 2300
Page 3: ...This manual covers the following Valhalla Scientific products Models 2300 2301 2300L and 2301L...
Page 19: ......
Page 20: ......
Page 23: ......
Page 24: ......
Page 26: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 50: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 74: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 100: ......
Page 101: ......
Page 103: ......
Page 106: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 113: ......
Page 114: ......
Page 115: ......