6.3 Calibration
The following procedures should be performed at routine intervals to ensure that the Power
Analyzer remains within specified limits. In addition, calibration should be performed following
repairs involving accuracy determining components.
Remove the top cover from the unit to gain access to the internal adjustments. The locations of
the adjustments are listed on the circuit board and on drawing number 2105-600 at the back of
this manual. Apply power to the 2100 and test equipment and allow approximately 30 minutes
for stabilization.
Note that the instrument may also be returned to the factory for full calibration traceable to
NIST.
6.3.1 Equipment Required
Qty 1 – Multi Calibrator (Fluke 5500A or equivalent)
Qty 1 – Precision Multimeter DMM (Agilent 3458A or equivalent)
6.3.2
ADC Reference Adjustments
1.
Leave the DVM COMMON on the GND POST, and connect the DVM POSITIVE lead to
TP5.
2.
Adjust RV22 for a 1.0000V on the DVM.
3.
Connect the positive lead to TP6.
4.
Adjust RV23 for a 1.0000V on the DVM.
5.
DVM Negative lead to TP7, DVM Positive lead TP8. Apply +10A
Adjust RV1 for -100.00mV. (see figure 6-1 for connections to calibrator)
6.
DVM POSITIVE lead to TP1.
7.
Close calibrator lid.