8
3.
Specifications
This section contains accuracy and operational specifications for the Model 4314Ki. The accuracy
specifications are valid for a period of one year from the date of calibration at a temperature of 15
°
C
(59
°
F) to 35
°
C (95
°
F). Outside this range, the temperature coefficient specification applies.
Range/Resolution/Test Current/Failsafe Current
Table 1
Range
Value
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
2MΩ
Standby
Resolution
100μΩ
1mΩ
10mΩ 100mΩ
1Ω
10Ω
100Ω
1kΩ
-
Test
Current
5mA
5mA
500µA
50μA
5μA
500nA
50nA
5nA
0A
Fail-Safe
Current
[1]
8mA
8mA
1.8mA 180μA
18μA
1.8μA
180nA
18nA
-
[1] Maximum value. Actual Failsafe Currents vary from instrument to instrument. The actual measured
level is provided on the certificate of calibration.
Accuracy
Table 2
Range
Accuracy
2Ω
±0.03% of reading ± 0.02% of range
20Ω
±0.03% of reading ± 0.02% of range
200Ω
±0.03% of reading ± 0.02% of range
2kΩ
±0.03% of reading ± 0.02% of range
20kΩ
±0.03% of reading ± 0.02% of range
200kΩ
±0.05% of reading ± 0.05% of range
2MΩ
±1% of reading ± 0.2% of range
20MΩ
±2% of reading ± 0.2% of range