25
LogOn2
COMMAND
Turns on continuous reading mode synonymous with “Screen? + RangeV?”.
Note: if the device is already Logging, the LogOn2 command will return
“Invalid Command”
Syntax:
LogOn2<crlf>
Returns:
Screen data + the value of the range every 500ms.
Example for 2
Ω
Range:
1.2345<crlf>2
Example for 20
Ω
Range:
12.345<crlf>20
Example for 200
Ω
Range:
123.45<crlf>200
Example for 2k
Ω
Range:
1.2345<crlf>2k
Example for 20k
Ω
Range:
12.345<crlf>20k
Example for 200k
Ω
Range:
123.45<crlf>200k
Example for 2M
Ω
Range:
1.2345<crlf>2M
LogOff
COMMAND
Stops continuous measurement mode.
Syntax:
LogOff<crlf>
Response: <crlf>
Ohms?
QUERY
Responds with the front panel reading formatted in ohms with multiplier.
No leading zeros.
Syntax:
Ohms?<crlf>
Returns:
Display data in Ohms with multiplier.
Example for a 1.1 ohm reading on:
2
Ω
Range:
1.1000
20
Ω
Range:
1.100
200
Ω
Range:
1.10
2k
Ω
Range:
0.0011k
RangeN?
QUERY
Responds with the range number from 1 to 8
Syntax:
RangeN?<crlf>
Returns:
R# (1-8)
R1
-
2 Ohms Range
R2
-
20 Ohms Range
R3
-
200 Ohms Range
R4
-
2k Ohms Range
R5
-
20k Ohms Range
R6
-
200k Ohms Range
R7
-
2M Ohms Range
R8
-
20M Ohms Range