5.3
Warning Message
If a Pressure Module is not fitted and the uvSVX is deployed, damage can be caused to the instrument. In
order to bring this situation to your notice a warning is transmitted on power up and putting the uvSVX into
run mode: #028 or the watch dog.
This message can be turned on or off
Code
Description
#220;n
Enable | Disable the Warning Message
#220;0 disables the warning
#120;1 enables the warning
#221
Reads the Warning Message activation status:
5.4
Error Flag
Code
Description
#035;nnnnn
Sets the value output in the output string when the Pressure | Depth value is in error
e.g. missing Pressure Module
Where n = a number from -100 000 to 100 000
Example:
#035;-9999
#036
Read the Error Flag
DataLog x2 showing a warning message that a Pressure Module is not fitted