background image

27

Advanced Theory of Operation

Dynacalibrator operation is based on the principle of mixing a known mass flow 
of gas derived from a permeation device with a metered stream of clean carrier/
dilution gas to generate a precise concentration of span gas.

User-selectable calibration gases originate from Dynacal permeation devices 
which are maintained at a user-selectable constant temperature within the 
Dynacalibrator’s permeation chamber.  Permeation rates can be precisely 
determined gravimetrically as a function of temperature.

An internal flow regulating system ensures a stable gas flow over extended 
periods of time for both the carrier air and dilution air streams.  The carrier stream 
continuously sweeps the permeation gas from the permeation chamber.  This 
stream then mixes with the dilution stream to provide the final calibration 
as supply.

Different span gas concentrations can be generated by changing one or a 
combination of the following parameters: (1) the dilution stream flow rate, (2) the 
permeation chamber temperature, or (3) the number and/or type of permeation 
devices.  The fastest, simplest, and most common method is to change only the 
dilution flow rate.

The carrier flows at a fixed rate, maintained at a constant value by a sapphire 
orifice plate and a differential pressure regulator.  Orifice sizes are available for 
rates of 80-1300 cc/min.

The dilution flow is user-adjustable from approximately 100 cc/min to over 
20 liters/min, depending on the capacity option selected.  Model 340 Dynacali-
brators contain a single dilution flow control valve and rotameter.  Since the 
Model 340 can provide a wide dynamic range of span gas calibrations simply 
by changing the flow rate of the dilution gas, it is considered to be a multi-point 
calibrator of exceptional range, accuracy, and precision.

Each Dynacalibrator model contains basic pneumatic plumbing, with several 
configuration options.  The plumbing configuration selected (see the simple dia-
gram below) processes the input gas supply into a carrier stream and a dilution 
stream. The carrier stream picks up the calibrated trace gas and is subsequently 
mixed with the dilution stream for the desired output concentration.

©

Figure 9:  Basic block diagram

Plumbing Configurations

There are six different plumbing configurations available for each of the  
Dynacalibrator models.  The plumbing configuration is defined by the first digit 
of the 2-digit numerical suffix to the basic model number, e.g., Model 340-12 is 
equipped with plumbing configuration 1.  The sec ond digit of the 2-digit suffix 
defines the overall instrument gas flow capacity.  A lower case letter is added to 

Summary of Contents for 340-0X

Page 1: ...es 713 688 9345 tech 713 688 8106 fax valco vici com VICI AG International Schenkon Switzerland Int 41 41 925 6200 phone Int 41 41 925 6201 fax info vici ch North America South America and Australia O...

Page 2: ...Attention Radiant heat warning Entanglement warning Electrical warning Symbols used in this document...

Page 3: ...icators 12 Main Power Switch 12 Heater Power Switch 12 Permeation Chamber 12 Permeation Chamber Flowmeter 12 Temperature Readout and Controller 12 Dilution Flow Control Valves 12 Dilution Flowmeters 1...

Page 4: ...nspection and Cleaning 33 Carrier Flow Rate Verification 34 Dilution Flowmeter Calibration 34 Chamber Temperature Calibration 35 Troubleshooting 36 Electrical Troubleshooting 36 Pneumatic Troubleshoot...

Page 5: ...rent compounds both organic and inorganic How to UseThis Manual This manual provides installation operation and maintenance information for all configurations of the Model 340 The identification tag o...

Page 6: ...y DPR1 The output of the permeation chamber is fed to a mixing tee where the calibrat ed dilution stream is added to set the final concentration The combined stream is then fed through another tee to...

Page 7: ...ibration It can be remotely switched between the zero air and span output modes A front panel switch permits selection of the ZERO SPAN or REMOTE mode In the ZERO mode a two position valve directs the...

Page 8: ...to carrier and dilution streams The gas stream bypasses on these units are also downstream from the pump permitting the user to connect external scrubbers driers separate gas supplies etc in either or...

Page 9: ...5 Specifications Introduction...

Page 10: ...laim Unpacking the Dynacalibrator After the initial receiving inspection and check the Dynacalibrator may be unpacked A carefully sized and mated pair of shipping cartons have been designed to ensure...

Page 11: ...ent to VICI for inspection and repair When this work is completed the equipment will be returned to you with an invoice itemizing charges for all repairs This invoice will become part of your claim to...

Page 12: ...alibrated A rack mounted enclosure in a mobile installation van aircraft etc requires additional mechanical support between the rack and the rear of the enclosure This addi tional support is not neede...

Page 13: ...340 5X O O O X X X O O 340 0X Z X X X O O 340 1X Z X X X X O O 340 2X Z O X X O O 340 3X Z O O X X X O O 340 4X Z O O X X X O O 340 5X Z O O O X X X X O O X Required connection O Optional connection b...

Page 14: ...alibrator is to be operated in an environment that has signifi cant concentrations of dust or other particulate matter we recommend a VICI Metronics Particulate Filter Assembly PFA connected to the su...

Page 15: ...e either perma nently plumbed to the calibrate input port of an analyzer or temporarily connected to its sample input for calibration depending on the capabilities of the analyzer NOTE The span outlet...

Page 16: ...e factory Refer to the calibration sheet at the back of the manual TEMPERATURE READOUT AND CONTROLLER The front panel displays the current chamber temperature and the control status of the instrument...

Page 17: ...tch selects the calibration mode LEDs indicate the current mode ZERO The permeation chamber output is dumped out the chamber vent only the dilution stream exits through the stream outlet to the analyz...

Page 18: ...34 T TO Where PO Permeation rate at reference temperature TO P New permeation rate at temperature T Certified devices should be used only at the temperature specified on the certificate Using any perm...

Page 19: ...the span outlet to your instrument Always leak check the entire instrument and all flow connec tions particularly if toxic corrosive or flammable gas mixes will be generated Use clean dry air or N2 to...

Page 20: ...and revert to its previous status To prevent the set point from being accidentally changed the front panel controls can be disabled with a serial command Refer to the section entitled Serial Port Com...

Page 21: ...leak tight 8 Reinstall the front panel assembly and rotate the panel lock screw 90 clockwise 9 Turn on the heater power switch and allow one hour for equilibration Refer to the separate instructions a...

Page 22: ...t is equipped with an internal pump the floats in both flowmeters will rise If the unit does not have an internal pump start the external supply and note that the floats rise The pressure of the exter...

Page 23: ...r illuminates 15 Let the unit sit under power until the temperature equilibration indicator flashes regularly about once every second This usually takes approximately 1 5 hours 16 After the chamber te...

Page 24: ...r is off and restored at power up Command Meaning DC n Show set decimation count Data logging rate in seconds L Start data logging Log data at the interval set by the decimation count DC 1 sets a one...

Page 25: ...ssary The following screen will appear 3 Click OK which reveals the next screen 4 Enter a name such as dyna340 and select an icon to associate with the file Then click OK 5 If necessary click through...

Page 26: ...opriate COM port and click OK to bring up this screen 7 Set the port parameters as indicated above and click OK 8 Now we are at the dyna340 HyperTerminal screen below Select File then Properties Seria...

Page 27: ...23 9 On the dyna340 Properties screen above click Settings to bring up the screen below Click Settings Click ASCII Setup Serial Port Communication...

Page 28: ...izing IIC DS1624 started Valco Dynacalibrator online A simple processor test can be performed by opening HyperTerminal before powering up the Dynacalibrator If the processor has some defects the messa...

Page 29: ...alone exceeds the suction rate of all the analyzer s being calibrated 3 The usual flow rate setting is the one which was calculated for the first span point Align the center of the float in the flowm...

Page 30: ...4 Unscrew the chamber cap with the tool provided 5 Add or remove the permeation device s with the supplied forceps or other tool appropriate to the job 6 Secure the chamber cap with the tool provided...

Page 31: ...d at a constant value by a sapphire orifice plate and a differential pressure regulator Orifice sizes are available for rates of 80 1300 cc min The dilution flow is user adjustable from approximately...

Page 32: ...tee Dry gas air nitrogen etc at the appropriate pressure is supplied by the user to the supply inlet on the back panel Particulates larger than 30 microns are removed by filter F1 Line regulator PR1 i...

Page 33: ...nel which must be plugged if an external carrier supply is used This port can be connected through external scrubbing media filters driers etc or simply looped to the carrier inlet port Filter F2 remo...

Page 34: ...30 AdvancedTheory of Operation Figure 8 Diagram of Configuration 3 Figure 9 Diagram of Configuration 4...

Page 35: ...e RV1 set at 5 psig The tee at the output of the accumulator splits the supply into carrier and dilution streams each of which is plumbed to a back panel outlet port which must be plugged if an extern...

Page 36: ...24 VDC If the cord end type is inappropriate for your locale please contact VICI for a replacement cord The main power switch S1 is on the front panel along with the main circuit breaker CB1 When the...

Page 37: ...4000 hours Instrument Inspection and Cleaning The following procedure should be performed quarterly for units in continuous service or at least after every 1000 hours of service For the steps which ca...

Page 38: ...hanges in its reading can indicate the need for flow verification To verify the flow rate a one liter N I S T traceable flowmeter is recommended After using the appropriate leak checking proce dures t...

Page 39: ...n modeled and corrections are calculated based on the model response The corrections are input and then the multiple setpoints are again run with the PRT and must fall within the specification in orde...

Page 40: ...al shock and serious injury to personnel PneumaticTroubleshooting There are two reasonably direct methods that can be used to troubleshoot leaks in a Dynacalibrator each with its advantages and disadv...

Page 41: ...allen screws on each side of the instrument above the side bars 2 Trace out the internal plumbing loop that has been found to be leaking starting at the point where the leak checker gauge has been co...

Page 42: ...er Pneumatic Issues Three other general categories of problems aside from leaks are occasionally encountered in the pneumatic system in the Dynacalibrator In order of likelihood they are Main pump dia...

Page 43: ...39 Model Number Breakdown...

Page 44: ...om or relating in any manner to these goods This Limited Warranty does not cover lamps or fuses damage due to improper shipping damage due to improper use damage due to modifications or alterations da...