Touchscreen display and main views (page 31)
‣
Starting up transmitter using touchscreen display (page 34)
2.5 Web interface
The transmitter also has a web interface for remote access. You can configure the web
interface to show numeric readings and graphs of 1 … 6 measurement parameters at the same
time. You can also configure the transmitter outputs and other transmitter settings, and
update the transmitter software.
The web interface has 2 user levels:
•
Administrator
: Configuration rights. Can configure outputs and change transmitter
settings.
•
Guest
: View-only access. Can add and remove measurement parameters in the
Measurements
view.
The web interface supports most major browsers (for example, Firefox, Chrome, and Safari):
using the most recent version is recommended.
More information
‣
Web interface and main views (page 33)
‣
Starting up transmitter using web interface (page 38)
‣
Updating transmitter software version (page 64)
2.6 Output options
The transmitter provides 4 analog output channels, 2 relays, and an Ethernet connection for
the Modbus TCP/IP protocol and web interface.
If the transmitter is powered using Power over Ethernet (PoE), analog outputs
and relays are not available.
2.6.1 Analog outputs
The transmitter provides 4 scalable analog output channels with voltage or current output.
Available analog output modes:
• Voltage: 0 … 1 V, 0 … 5 V, 0 … 10 V
• Current: 4 … 20 mA, 0 … 20 mA
Use the touchscreen or web interface to configure the output mode for the channels (same
output mode in all channels), as well as the measurement parameter and scaling for each
channel.
Indigo 520 User Guide
M212287EN-A
12