Adjusting Humidity and Temperature with
Push Buttons
The instructions here apply to both 1-point and 2-point calibration.
To make a humidity or temperature adjustment:
1. Open the transmitter cover. There are 3 buttons marked
ADJ
,
+
and
-
on the motherboard.
There are also two indicator LEDs, one green and one red.
2. Press the
ADJ
button and hold it down until the green indicator LED begins to blink slowly
(800ms cycle time).
3. The transmitter is now in RH calibration state. Analog output and optional display will still
follow the actual measured RH value.
If you do not wish to perform the RH adjustment at this time, press the
ADJ
button one more time. The green indicator LED turns off and the red
indicator LED begins to blink slowly (800ms cycle time) to indicate T
calibration state. To calibrate T, follow the instructions starting from
.
4. Remove the filter and insert the probe into a measurement hole of the dry end reference
chamber (for example, LiCl: 11 % RH) to do the low humidity offset adjustment.
Do not touch the adjustment buttons before the conditions have stabilized.
This takes approximately 30 minutes.
5. Press either the
-
or
+
button at least once, make sure that the I
out
current is correct,
adjust using the
-
and
+
buttons if needed, and press the
ADJ
button again. The green
indicator LED begins to blink faster (400ms cycle time).
If you do not wish to perform the two-point RH adjustment at this time, press
the
ADJ
button one more time. The green indicator LED turns off and the red
indicator LED begins to blink slowly (800ms cycle time) to indicate T
calibration state. To calibrate T, follow the instructions starting from
.
If one-point calibration is done at more than 50% RH, a gain adjustment is
done instead of an offset adjustment.
6. Insert the probe into the high end reference chamber (for example, NaCl: 75 % RH
chamber in the humidity calibrator HMK15) and do the high humidity gain adjustment by
using the - and + buttons to make sure the I
out
current is correct (you have to press either
-
or
+
at least once even if the value is correct). To finish the RH calibration, press the
ADJ
button. The green LED is now turned off and the red indicator LED begins to blink slowly
(800ms cycle time).
10
M211247EN-E
Summary of Contents for HUMICAP HMT120 Series
Page 4: ...4 M211247EN E...
Page 20: ...Kanalinstallationssatz 20 M211247EN E...
Page 32: ...Kit d installation sur gaine 32 M211247EN E...
Page 44: ...Kit de instalaci n de conductos 44 M211247EN E...
Page 52: ...52 M211247EN E...
Page 56: ...Kit de instala o de duto 56 M211247EN E...
Page 64: ...64 M211247EN E...
Page 65: ...Vaisala HUMICAP HMT120 Td RH T HMT120 3 5 10 20 m LCD HMT120 4 65...
Page 66: ...145 72 120 37 124 12 116 46 39 M4 4 91 92 78 66 M211247EN E...
Page 67: ...1 2 67...
Page 75: ...www vaisala com warranty helpdesk vaisala com www vaisala com support 75...
Page 76: ...76 M211247EN E...
Page 77: ...Vaisala HUMICAP HMT120 Td RH T HMT120 3 5 10 20 HMT120 LCD HMT120 77...
Page 78: ...145 72 120 37 124 12 116 46 39 M4 4 91 92 78 78 M211247EN E...
Page 79: ...1 2 79...
Page 87: ...www vaisala com warranty Vaisala helpdesk vaisala com www vaisala com support 87...
Page 88: ...88 M211247EN E...
Page 89: ...Vaisala HUMICAP HMT120 Td RH T HMT120 3 5 10 20 HMT120 HMT120 89...
Page 90: ...145 72 120 37 124 12 116 46 39 M4 4 91 92 78 90 M211247EN E...
Page 91: ...1 2 91...
Page 95: ...7 HMT120 T T T ADJ 8 HMK15 9 ADJ 400 T ADJ 10 11 ADJ 12 ADJ 200 2 95...
Page 96: ...HMT130 Probe cal RH 1 Probe cal RH 2 Probe cal T1 Probe cal T2 Probe cal Error 96 M211247EN E...
Page 100: ...www vaisala com warranty Vaisala helpdesk vaisala com www vaisala com support 100 M211247EN E...
Page 101: ......
Page 102: ...Drilling Template 1 6 5 4 2 46 39 91 92 78 56 36 5 5...
Page 103: ...www vaisala com...