Property
Description/value
Accuracy at +23 °C (+73.4 °F)
±0.8 %RH (0 … 90 %RH)
Factory calibration uncertainty
±0.7 %RH (0 … 40 %RH)
±1 %RH (40 … 95 %RH)
Non-linearity
0.4 %RH
T
63
response time
15 s
Sensor
HUMICAP
â
I
Temperature
Measurement range
−40 … +120 °C (−40 … +248 °F)
Accuracy at +23 °C (+73.4 °F)
±0.1 °C (±0.18 °F)
Factory calibration uncertainty
±0.1 °C (±0.18 °F) at +23 °C (+73.4 °F)
T
63
response time
70 s
1)
Defined against calibration reference. Including non-linearity, hysteresis, and repeatability.
2) Defined as ±2 standard deviation limits. Small variations possible; see calibration certificate.
3) In still air
Table 27 Operating environment
Property
Description/value
Operating temperature of probe body
−40 … +60 °C (−40 … +140 °F)
Operating temperature of probe head
−40 … +120 °C (−40 … +248 °F)
Storage temperature
−40 … +60 °C (−40 … +140 °F)
Operating environment
Suitable for outdoor use when protected from
rain
Measurement environment
For air, nitrogen, hydrogen, argon, helium, and
oxygen
IP rating
IP65 for probe body
EMC compliance
EN 61326-1, industrial electromagnetic
environment
1)
Consult Vaisala if other chemicals are present. Consider safety regulations with flammable
gases.
Chapter 8 – Technical data
65