USER'S GUIDE ____________________________________________________________________
132 __________________________________________________________________ M211187EN-C
If any of the tests fails, it can be repeated up to two times. If any of the
tests finally fails, the visibility calibration of the measurement unit is
reported as invalid and the data values are invalidated.
Once the Guided Calibration Procedure has started it has to be completed
either as passing or as failing. The procedure cannot be aborted or
suspended. At the end of the Guided Calibration Procedure a
summarizing report is generated.
1.
Give the
CALIBRATE FS11
command.
2.
The instrument responses:
Step 1/3
Clean Windows and type YES:
YES
........||||||||||
3.
Clean the measurement unit windows.
- Use a soft, lint-free cloth and a suitable window cleaning
detergent to clean the windows.
- Use a soft, lint-free and dry cloth to remove remaining detergent
residues entirely.
- Be careful not to scratch the window surfaces.
4.
Type YES and press enter.
The test step can take up to two minutes.
5.
The instrument responses:
Step 2/3
Install Zero Plugs and type YES:
YES
6.
Install the zero plugs to the measurement unit transmitter and
receiver. See Figure 40 on page 128.
7.
Type YES and press enter.
- The Zero Signal Test starts and the progress is indicated on the
screen.
- When the Zero Signal Test is finalized a pass/fail indication is
given and the resulting Zero Signal is given in brackets.
- The test step can take up to two minutes.
Step 2/3
Install Zero Plugs and type YES:
YES
........||||||||||
Zero signal test PASSED. (0.0000)