Syntax
Description
FORM [x]
<cr>
[x]
= formatter string
The formatter string consists of parameters and modifiers. The
maximum length of the form string is 74 characters. When
entering the command, use the abbreviations of the parameters.
The parameters are presented in
modifiers in
.
FORM /
<cr>
Return to default format.
Example:
form 4.2 TDF #r #n
OK
form 3.1 “H2O= “ ppm “ “ U3 #r #n
OK
Table 18 FORM Command Parameters
Parameter
Description
TDF
Dew point / frost point temperature
PPM
Water content in parts per million
PPB
Water content in parts per billion
PPMW
Water content in parts per million by weight
For accurate ppm, ppb, or ppm
w
reading, you must compensate for the pressure
and molar mass of the process gas.
Table 19 FORM Command Modifiers
Modifier
Description
x.y
Length modifier (number of digits and decimal places)
#t
Tabulator
#r
Carriage return
#n
Line feed
""
String constant
#xxx
ASCII code value (decimal) of a special character; for example,
#027
for
Esc