![Vaisala Drycap DMT152 User Manual Download Page 36](http://html1.mh-extra.com/html/vaisala/drycap-dmt152/drycap-dmt152_user-manual_3228362036.webp)
The final status character can have several values:
0
Normal status
1
Transmitter error
7
Retrying auto-calibration due to fluctuations in the supply voltage
9
Transmitter startup
Some example status codes:
0109
Transmitter performing startup procedure
1100
Sensor purge in progress, output frozen
1110
Settling period after purge, output frozen
0000
Normal measurement
Table 10 STAT command
Syntax
Description
STAT
<cr>
Display the status of the transmitter in 4
characters.
Example:
STAT
0000
It is normal for measurement output to freeze temporarily if the measurement
environment is changing rapidly.
5.2.3 Show Firmware Version
Table 11 VERS command
Syntax
Description
VERS
<cr>
Display the transmitter firmware version
number.
Example:
vers
DMT152 / 1.10
DMT152 User Guide
M210865EN-J
34
Summary of Contents for Drycap DMT152
Page 1: ...M210865EN J User Guide Vaisala DRYCAP Dew Point Transmitter DMT152...
Page 8: ...DMT152 User Guide M210865EN J 6...
Page 68: ...DMT152 User Guide M210865EN J 66...
Page 70: ...DMT152 User Guide M210865EN J 68...
Page 71: ......
Page 72: ...www vaisala com...