USER'S GUIDE____________________________________________________________________
46 __________________________________________________________________ M211371EN-B
Set Serial Line Response Time
With the
SDELAY
command you can set delay (response time) of the
serial line, or view the currently set delay value. Increasing the value
by one increases the delay by 4 milliseconds. For example, the default
value is 10 which equals a ~40 ms answer delay.
SDELAY
[
delay
]<cr>
where
delay = Serial line delay, range 0 … 255.
Example:
sdelay 20
Serial delay : 20
Other Commands
Opening the Transmitter in POLL Mode
Use the
OPEN
command to connect to a transmitter that is in POLL
mode.
OPEN
[
address
]<cr>
where
address = Transmitter address, range 0 ... 255.
Example (target transmitter in POLL mode, with address 5):
open 5
DPT145 5 line opened for operator commands