• 2-point adjustment is recommended if you typically measure a variable CO

2

 level.

Available adjustment functions depend on the interface you use to operate the GMP231. If you
want to integrate the functionality into the incubator’s control system, the I

2

C interface and the

Vaisala Industrial Protocol are recommended. If you want to compare the reading of the
GMP231 to a reference instrument as adjust it accordingly, use an MI70 handheld indicator and

a reference probe.

6.4.4.1 I

2

C Interface

I

2

C interface supports 1 and 2-point adjustment sequences as well as configuration of the

environmental compensation settings using the Get_Parameter and Set_Parameter
commands.

More information

Get_Parameter (page 55)

Set_Parameter (page 57)

Adjust (page 61)

6.4.4.2 Vaisala Industrial Protocol

Vaisala industrial protocol supports one and two-point adjustment with the 

CCO2

 command.

You can also adjust the internal pressure and temperature measurements with a simple offset
correction. Configuration of the environmental compensation settings can be done using serial
line commands.

More information

Changing measurement settings (page 35)

Calibration commands (page 40)

6.4.4.3 MI70 handheld indicator

MI70 handheld indicator supports 1-point adjustment, either using a calibration gas or using a

reference instrument that is connected to the MI70.

More information

Calibration and adjustment with MI70 indicator (page 77)

GMP231 User Guide

M211501EN-E

74

Summary of Contents for CARBOCAP GMP231

Page 1: ...M211501EN E User Guide Carbon dioxide probe for CO2 incubators GMP231...

Page 2: ...ange without prior notice Local rules and regulations may vary and they shall take precedence over the information contained in this document Vaisala makes no representations on this document s compli...

Page 3: ...atory compliances 16 3 Installation 17 3 1 Thermal management 17 3 2 Avoiding condensation 17 3 3 Probe installation depth 17 3 4 Dimensions 18 3 5 Recommended installation 20 3 6 Wiring 22 3 7 Power...

Page 4: ...anging the filter 69 6 3 Changing the silicone plug 70 6 4 Calibration and adjustment 71 6 4 1 Calibration setup 71 6 4 2 Effect of environmental compensations 73 6 4 3 Limits of adjustment 73 6 4 4 A...

Page 5: ...Online store 87 Warranty 87 Technical support 87 Recycling 87 Table of contents 3...

Page 6: ...igure 8 Electronics housing dimensions 19 Figure 9 Attachment bracket dimensions 20 Figure 10 Recommended installation 21 Figure 11 PuTTY terminal application 27 Figure 12 Hardware schematic 50 Figure...

Page 7: ...3 Table 20 Output commands for FORM command 34 Table 21 Modifiers for FORM command 35 Table 22 Compensation mode commands 35 Table 23 ENV command 36 Table 24 STANDBY command 38 Table 25 SMODE command...

Page 8: ...ure sequence example 59 Table 59 Get_Parameter_Info invoke message 60 Table 60 Get_Parameter_Info response message 60 Table 61 Parameter data types 60 Table 62 Parameter persistence 61 Table 63 Adjust...

Page 9: ...1501EN D December 2014 Applicable from software version 1 1 0 onward Updated section Status word page 52 M211501EN C May 2014 Updated description for ENV command in section Changing measurement settin...

Page 10: ...mentioned in this publication are trade names trademarks or registered trademarks of their respective owners 1 5 Patent notice This product is protected by the following patents and their correspondi...

Page 11: ...eat the chamber interior to 180 C Special design of the probe enables easy installation through the incubator wall and ensures that the sensor optics and electronics are separated and thermally insula...

Page 12: ...temperature and background gas Pressure compensation based on integrated pressure sensor Heating to avoid condensation on optical elements Analog output one current output channel 0 20 mA or 4 20 mA D...

Page 13: ...10649 M4 3 Type label on housing cover not shown 4 8 pin M12 connector 5 Probe body 6 Measurement cuvette with optics and CARBOCAP CO2 sensor 7 PTFE filter 8 Silicone plug for sealing the lead throug...

Page 14: ...ith a thermopile chip at a wavelength chosen by a micromechanical Fabry P rot interferometer FPI and a band pass filter A hermetically sealed sapphire window is used to protect the sensor chips from m...

Page 15: ...nvironmental compensation GMP231 improves the CO2 measurement accuracy by applying various environmental compensations GMP231 compensates for the effects of Pressure Temperature Background gas oxygen...

Page 16: ...ation and is recommended for use unless a dedicated temperature measurement is available and can be regularly updated to the GMP231 If temperature compensation is turned off GMP231 uses the default va...

Page 17: ...the error state which is 23 mA for the 0 20 mA output This overrange and error state behavior is specific to the analog output and does not affect the readings provided by the digital outputs You can...

Page 18: ...are unable to take either precaution touch a conductive part of the equipment chassis with your other hand before touching ESD sensitive components Hold component boards by the edges and avoid touchin...

Page 19: ...stallation tube must be sealed from the chamber side to limit heat conduction and to prevent CO2 in the chamber from entering the probe Vaisala recommends a 44 mm diameter installation tube together w...

Page 20: ...4 Dimensions 163 118 5 Screw BN 10649 M4 x 8 mm for 1 mm thick sheet metal attachment 11 21 30 2 1 3 19 24 8 Position only the filter inside the heated chamber Figure 6 Probe dimensions without silic...

Page 21: ...19 2 5 Incubator s inner wall 1 3 Position only the filter inside the heated chamber Recommended diameter for installation tube 44 mm Figure 7 Probe dimensions with silicone plug M12 54 4 87 Reserve s...

Page 22: ...hment bracket and the two screw holes on the side of the probe As the installation depth of the probe inside the chamber is critical the mounting method must allow the probe to be positioned exactly T...

Page 23: ...obe housing 6 8 pin M12 connector 7 Cable 8 CARBOCAP sensor under PTFE filter The filter must be inside the chamber completely 9 Silicone plug Insert the plug over the probe from inside the chamber 10...

Page 24: ...pin 5 is no longer grounded unless its internal measurement shows the temperature is too high Table 4 GMP231 connector pinout Male 8 pin M12 Pin Function 1 5 6 7 3 4 8 2 1 I2C SDA 2 RS 485 D 3 I2C SCL...

Page 25: ...2C SCL Green 4 Analog output Yellow 5 Standby Gray 6 RS 485 D Pink 7 Power supply Blue 8 Ground Red Shield Black 3 7 Power supply The supply voltage range of GMP231 is 11 30 VDC If the analog output i...

Page 26: ...terface settings Property Value Bit rate 19200 Parity None Data bits 8 Stop bit 1 Flow control None 4 3 Physical interface The physical interface is a non isolated 3 wire interface The data lines are...

Page 27: ...es If the device is listed as Vaisala USB Device with a COM port number in brackets the cable is ready for use Note the COM port number you will need it later If the device is listed as Vaisala USB In...

Page 28: ...or USB line to connect to field If you are using the PuTTY terminal application supplied by Vaisala you can press the USB Finder button to open the Vaisala USB Instrument Finder program 4 Check that t...

Page 29: ...commands All commands can be issued either in uppercase or lowercase In the command examples the keyboard input by the user is in bold The notation cr refers to pressing the carriage return ENTER key...

Page 30: ...r set the serial interface settings SMODE mode Show or set startup serial mode RUN STOP or POLL SNUM Show probe serial number STANDBY on off Standby mode on off turn off measurement SYSTEM Show probe...

Page 31: ...ription cr Show listing of device information cr Show listing of device information even if the device is in POLL mode but its address is unknown Do not use the command if you have more than one devic...

Page 32: ...scription TIME Show how long the probe has been in operation since the last startup or reset The operation counter is in format hh mm ss Full days and hours are retained in the counter minutes and sec...

Page 33: ...mand Syntax Description R cr Start the continuous outputting of measurement values as an ASCII text string to the serial line The probe keeps outputting measurement messages at the interval that has b...

Page 34: ...measurement messages are repeated in the RUN mode or after the R command has been given Set the output interval n interval range 0 255 xxx unit for interval setting s seconds min minutes h hours If yo...

Page 35: ...urement format to default FORM modifier string cr modifier string String of parameters and modifiers that defines the output format length 1 150 characters Maximum length may be shorter when text stri...

Page 36: ...example CO2 3563 ppm 9F CO2 3562 ppm 9E CO2 3559 ppm A4 Command to set output format as CO2 ppm with start of text ASCII character 002 and end of text 003 ASCII codes and without line feed and carriag...

Page 37: ...ASCII code value decimal of a special character for example 027 for ESC ADDR Probe address 0 254 SN Probe serial number TIME Cumulative operating hours of the probe Ux Shows the name of the measureme...

Page 38: ...rement Available only for pressure and temperature compensations Example check temperature compensation mode pass 1300 tcmode T COMP MODE ON Example change temperature compensation mode to use interna...

Page 39: ...ation you must write all continuously updated values to RAM to avoid wearing out the EEPROM Example Show current compensation values all compensations are enabled Note that temperature and pressure co...

Page 40: ...standby mode standby on Standby mode ON 4 9 Configuring serial line operation Table 25 SMODE command Syntax Description SMODE mode cr Set serial line start up operating mode Available modes STOP No a...

Page 41: ...Com1 Baud rate 19200 Com1 Parity N Com1 Data bits 8 Com1 Stop bits 1 Example set serial line to 9600 baud even 7 data bits and 1 stop bit seri 9600 e 7 1 OK seri Com1 Baud rate 9600 Com1 Parity E Com1...

Page 42: ...librating GMP231 using serial line commands see Calibration and adjustment page 71 Make sure that the environmental compensation settings of the GMP231 are properly set for your calibration environmen...

Page 43: ...ts Example show current user adjustment status no adjustment done pass 1300 cco2 1 Ref point low 0 1 Meas point low 0 2 Ref point low 200000 2 Meas point low 200000 Gain 1 0000 Offset 0 0000 Example o...

Page 44: ...acters If you want to use spaces in the string enclose the text in quotes Example shows the text Calibrated at followed by the currently stored calibration text pass 1300 ctext Calibrated at Vaisala E...

Page 45: ...orm an offset correction to the internal pressure measurement pressure Correct pressure in hectopascals hPa CP reset cr Clear the offset adjustment from internal pressure measurement Example adjust th...

Page 46: ...Lower limit of channel scaling in ppm highlimit High limit of channel scaling in ppm Example change channel scaling to 0 5 pass 1300 asel 1 co2 0 50000 Aout 1 quantity CO2 0 50000 Table 36 AMODE comm...

Page 47: ...w the behavior of the analog output when the measured value is outside the scaled output range channel Analog output channel only 1 can be selected AOVER channel clip valid cr Set the behavior of the...

Page 48: ...ns scaled to 0 50000 ppm Error limit is 10 which means the output will show the error state 23 mA when the measured value is 10 outside the scaled output range With the settings above this will happen...

Page 49: ...s advanced serial commands You can use advanced commands until the next reset code Code for enabling advanced commands 1300 Example pass 1300 Table 40 OPEN command Syntax Description OPEN address cr C...

Page 50: ...of the I2C interface Example showing normal status no active errors errs NO CRITICAL ERRORS NO ERRORS NO WARNINGS STATUS NORMAL Example showing probe in standby mode no active errors errs NO CRITICAL...

Page 51: ...s factory configuration All user settings are lost After using the FRESTORE command reset the probe using the RESET command Example pass 1300 frestore Parameters restored to factory defaults reset GMP...

Page 52: ...with power supply There are small pull up resistors for SCL and SDA Connector is 8 pin male M12 Maximum cable length should not exceed 5 m and maximum capacitance between communication lines and grou...

Page 53: ...GMP231 Device Address 0 0 0 0 1 0 0 1 MSB Device type Sub address Make sure that there are no addressing conflicts if other I2C devices are put on the same bus The GMP231 address can be changed More i...

Page 54: ...lid the message is rejected and the internal state of the GMP231 is set to Idle When the response message is transmitted the checksum is calculated using bytes from the status byte to the last byte of...

Page 55: ...am memory corrupted CRC error Fatal error contact Vaisala 1 Critical Error Parameter memory corrupted CRC error Fatal error contact Vaisala 2 Reserved 3 Reserved 4 Error Low supply voltage error Check...

Page 56: ...e set by user command Set to normal mode when desired 26 Status CO2 adjustment state on Complete the CO2 adjustment 27 Reserved 28 Reserved 29 Reserved 30 Reserved 31 Reserved More information Status...

Page 57: ...d set version 1 byte Parameter set version Checksum 2 bytes See Checksum page 52 5 2 2 Get_Parameter Get_Parameter command ID 81hex reads parameter values You can use it to read both measurement resul...

Page 58: ...h 07hex 39hex Data 1 byte ID of parameter 1 50 bytes Value of parameter If requested parameter ID is unknown these data bytes are left out of the response Checksum 2 bytes See Checksum page 52 5 2 2 1...

Page 59: ...GMP231 may not change immediately The length of the invoke message varies depending on the data length The master must write as many bytes as indicated in the frame length of the invoke message If th...

Page 60: ...See Table 46 page 51 Frame length 1 byte Always 08hex Data 1 byte Parameter ID 1 byte Return code See the following table Checksum 2 bytes See Checksum page 52 Table 57 Set_Parameter return codes Cod...

Page 61: ...tion 13 I2C address read 04 GMP231 Status byte error active ACK 82 Set_parameter command 09 Device address 08 Frame length 10 Parameter ID of compensation temperature 00 Return code Ok D1 CRC high CE...

Page 62: ...age 51 Command ID 1 byte Always 83hex Device address 1 byte See Table 46 page 51 Frame length 1 byte Data 1 byte Parameter ID 1 byte Parameter datatype See the following table 1 byte Parameter length...

Page 63: ...ddress 1 byte See Table 46 page 51 Frame length 1 byte Varies according to data length 07hex 0Bhex Data 1 byte Adjustment subcommand See Table 65 page 62 1 byte Parameter to be adjusted See Table 66 p...

Page 64: ...eference value is given in parameter 4 Cancel adjustment previous adjustment is reverted 5 End adjustment is saved and new values are taken in use 6 Revert to factory calibration clear user adjustment...

Page 65: ...nvironment and wait until the measurement has stabilized 3 Using the Adjust command record the measured point and give the reference value 4 Use the Adjust command to end the adjustment 5 3 2 2 point...

Page 66: ...e 0 65535 Long integer 4 Value range 2 147 483 648 2 147 483 647 Unsigned long integer 4 Value range 0 4 294 967 295 Float 4 According to IEEE 754 Nan 7FC00000hex is returned if no value is available...

Page 67: ...e Day ADATED 7 1 Byte R Yes Example 29 General parameters and variables Device status word STATUS 8 4 Str R No See Status word page 52 CO2 measurement mode CO2_MODE 9 1 Byte R W Yes 0 Standby mode Mea...

Page 68: ...re T_COMP 16 4 Float R W No Celsius Measured temperature internal sensor T 17 4 Float R No Celsius Pressure compensation mode PC_MODE 18 1 Byte R W Yes 0 Compensation off 1 Use pressure setpoint value...

Page 69: ...O2 point 2 CO2_RP2 27 4 Float R W Yes ppm CO2 Measured CO2 point 2 CO2_MP2 28 4 Float R Yes ppm CO2 Reference temperature point 1 T_RP1 29 4 Float R W Yes Celsius 0 C Measured temperature point 1 T_MP...

Page 70: ...W Yes ppm CO2 Analog output scale high 20 mA A_HI 38 4 Float R W Yes ppm CO2 Other Persistent compensation values used if volatile values are not given Compensation temperature T_COMP2 39 4 Float R W...

Page 71: ...he optical surfaces of the sensor in any other manner CAUTION More information Changing the filter page 69 Changing the silicone plug page 70 6 1 1 Chemical tolerance GMP231 tolerates the following st...

Page 72: ...igns of contamination It is best to replace the filter also 1 Pull the white PTFE filter away from the sensor Rotating the filter slightly makes it easier to detach 2 Lift one edge of the plug and pul...

Page 73: ...that affect the result Performing an accurate calibration and adjustment takes some time and preparation Instead of doing it yourself contact Vaisala Calibration means comparing GMP231 to a known refe...

Page 74: ...1 point calibration with the background CO2 level as your reference You need a calibrated reference instrument to compare against for example a GM70 hand held meter with a calibrated GMP222 probe With...

Page 75: ...en free For example pure nitrogen gas is typically used as a convenient 0 ppm CO2 reference Remember to restore the normal compensation settings after completing calibration and adjustment If you are...

Page 76: ...using the Get_Parameter and Set_Parameter commands More information Get_Parameter page 55 Set_Parameter page 57 Adjust page 61 6 4 4 2 Vaisala Industrial Protocol Vaisala industrial protocol supports...

Page 77: ...1 can remain physically connected to the incubator 2 If the MI70 indicator is on turn it off 3 Connect GMP231 to the MI70 indicator using the GMP231 service cable for MI70 4 Turn on the MI70 indicator...

Page 78: ...ompensation settings from Main menu Settings Measurement settings Figure 18 GMP231 compensation settings on MI70 screen When you turn a compensation off GMP231 still shows a value for the correspondin...

Page 79: ...1 point adjustment with an MI70 compatible reference probe 1 Connect the GMP231 to Port I of the MI70 indicator 2 Connect the calibrated reference probe to Port II Make sure the reference probe is in...

Page 80: ...e measured CO2 readings and their difference on the screen Allow the measurement to stabilize To proceed with the adjustment press Adjust 10 Select To same as CO2 II 11 You are prompted if you really...

Page 81: ...t sequence from Main menu Functions Adjustments 5 MI70 notifies you that automatic power off is disabled during adjustment mode press OK to acknowledge 6 Select the CO2 parameter when prompted 7 You m...

Page 82: ...the adjustment is successful MI70 shows the text Adjustment Done after which you return to the adjustment mode At this point you can press Back and Exit to leave the adjustment mode The adjustment is...

Page 83: ...probe on RS 485 line Incorrect wiring Check that the RS 485 connection is wired correctly Power cycle or reset the probe and try again Probe in POLL mode with unknown address Issue the command to make...

Page 84: ...de the scaled output range The default error level depends on the output type For 4 20 mA output the default error level is 3 6 mA For 0 20 mA output the default error level is 23 mA The probe resumes...

Page 85: ...O2 12 20 CO2 0 4 CO2 Non linearity at 0 20 CO2 0 1 CO2 Temperature dependence With compensation at 3 12 CO2 20 60 C 0 1 CO2 Without compensation typical 0 4 of reading C Pressure dependence With compe...

Page 86: ...rage temperature 40 75 C 40 167 F Pressure compensated 500 1100 hPa Pressure operating 1500 hPa Humidity 0 100 non condensing Condensation prevention Sensor head heating when power on Chemical toleran...

Page 87: ...n diameter 24 8 mm 0 98 in Probe tube length 118 5 mm 4 67 in Sensor filter diameter 19 mm 0 75 in Sensor filter length 12 mm 0 47 in 9 2 Spare parts and accessories Table 75 Spare parts and accessori...

Page 88: ......

Page 89: ...licable supply contract or Conditions of Sale for details of the warranty for each product Technical support Contact Vaisala technical support at helpdesk vaisala com Provide at least the following su...

Page 90: ......

Page 91: ......

Page 92: ...www vaisala com...

Reviews: