Table 28 SDELAY command
Syntax
Description
SDELAY
<cr>
Show serial line transmission delay.
SDELAY
[delay]
<cr>
Set a new serial line transmission delay, range 0
… 255 (milliseconds)
Example (set delay to 50 milliseconds):
sdelay
50
Serial delay : 50
Table 29 ECHO command
Syntax
Description
ECHO
[on/off]
<cr>
Enable or disable echoing of characters received
by the probe.
Example:
echo
Echo : OFF
4.10 Calibration commands
Before calibrating GMP231 using serial line commands, see
Make sure that the environmental compensation settings of the GMP231 are
properly set for your calibration environment.
Table 30 CCO2 command
Syntax
Description
cco2
<cr>
Show current user adjustment status.
cco2 -lo [co2]<cr>
cco2
-hi [co2]
<cr>
Perform a 1-point (only either low or high
concentration) or 2-point (both low and high
concentrations) calibration and adjustment.
-lo
= Adjustment at low concentration (under 2
%CO
2
)
-hi
= Adjustment at high concentration (over 2
%CO
2
)
co2
= CO
2
concentration reference in ppm
GMP231 User Guide
M211501EN-E
40
Summary of Contents for CARBOCAP GMP231
Page 1: ...M211501EN E User Guide Carbon dioxide probe for CO2 incubators GMP231...
Page 5: ...Online store 87 Warranty 87 Technical support 87 Recycling 87 Table of contents 3...
Page 88: ......
Page 90: ......
Page 91: ......
Page 92: ...www vaisala com...