User's Guide ______________________________________________________________________
112 _________________________________________________________________ M210855EN-C
LCP1 [x/y] [z]
Example
>lcp1
1. Reading ? 981.2
1. Reference ? 980.0
2. Reading ? 1102.1
2. Reference ? 1100.0
>
MPCP1
Use the
MPCP1
command.
-
to perform multipoint correction for barometer module/module P1
-
to activate or deactivate the multipoint adjustment function
-
to enter new linear offset and offset/gain pressure corrections to the
transmitter
-
to edit existing linear offset and offset/gain pressure corrections.
The command operates like
LCP1
, but can take up to eight reading/
reference pairs. If there are three pressure barometer modules installed,
the commands
MCP1
,
MCP2
, and
MCP3
will be available. First
deactivate the previous corrections by using the
MPCP1
OFF
command. Precalibration of the transmitter then gives the required
corrections. When entering new multipoint corrections, always start at
the low-pressure end and then go up the pressure range. Key in the
multipoint corrections for each barometer module separately.
where
x
=
ON
y
=
OFF
z
=
?
NOTE
The new linear corrections will always cancel the previous corrections
as well as the valid date of calibration of the transmitter.