0000137936
rev. 6.1
2
Instruction Setting
Visualization:
The instrument can control many different measures, for shown every measure is possible to select three
different screens with the key Up and Down. In the default screen are shown the Chlorine measure, the
Redox measure, the temperature measure and the pH measure. In the second screen are shown the
resettable totalizer and the date from the last reset, if the date is not set the display will write “install”. In the
third screen there are the Flow rate measure and the value of the permanent totalizer.
Functions:
•
Calibration (Press Cal Key for 3 Seconds):
o
Select the calibration routine for: Redox, pH, Chlorine or Flow rate by Up or Down key.
o
Standard Routine calibration pH probe is 7 and 4 buffer solution and Redox 465 mV buffer
solution.
•
Press Cal and Set Key (both) for 5 Seconds and run Program Setup:
o
Program_Menu
(Press Enter to set the following Item)
§
Language_
(It’s possible to have 5 language EN, CZ, DE, HR)
§
Advanced_
(Technical Service Only
password protected
)
§
RS485 SETTINGS
(only if is present)
•
RS-485___ON
(Adjust value with enter and up or down key to OFF
(disabled) or ON (Enable) the RS485 function
•
DOA___OFF (
Adjust value with enter and up or down key to OFF
(disabled) or ON (Enable) the function of recognition
)
•
Address_____1
(Adjust value from 1 to 247 to assign the ID at the
device)
•
Parity____EVEN
(Adjust the value to No (disabled), Even or Odd to
monitor the data transmission
)
•
BaudRate__19200
(Change the value from 1200, 2400, 4800, 9600 or
19200 to choose the velocity data transmission)
§
Rx_Measure
•
setpoint___750_mv
(Adjust value with enter and up or down key) It’s
possible to adjust from 0 to 1200 mV value for Redox
•
sp_type___low
(Adjust value LOW or HIGH)
•
ofa_time_000_min
(Change the value from 1 to 240 minutes or Off)
•
alr_band__000_mV
(Adjust value from 100 to 300 mV)
•
Type__PROP
(Adjust value between OFF, PROP, ON/OFF or TIMED)
(only if set to Timed there are two new entries to make dosage cycles ON /
OFF selectable)
o
Timed ON
(Change the value from 5 to 3600 seconds)
o
Timed OFF
(Change the value from 5 to 3600 seconds)
§
ph_Measure
•
setpoint____7.4ph
(Adjust value with enter and up or down key) It’s
possible to adjust from 0 to 14 pH value.
•
sp_type___acid
(Adjust value ACID or ALKA)
•
ofa_time_000_min
(Change the value from 1 to 240 minutes or Off)
•
alr_band__000_ph
(Adjust value from 1 pH to 3 pH)
•
Temp__25*C_
(Adjust value with enter and up or down key) pH measure
only.
enter
cal
set
esc
enter
cal
prog